background image

 

- 7 - 

- 8-

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

L/min

 

Mode No.6 Set the interval of updating the instantaneous flow rate display 

 

If the L/min switch is pressed when Mode No. 5 is lit, the screen will switch to Mode No. 6 and display the following. 
("

" shows the present setting value.) 

 
 
 
 
1.

 

When the L/min switch is pressed for 2 seconds, the setting value will blink, showing ready for setting. 

2.  Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is 0.5, 1.0, 2.0, 3.0, 4.0, 

or 5.0 sec.) 

(Default is 2.0 sec) 

3.  When the L/min switch is pressed for 2 seconds, the set value will be lit and setting will be completed. 

 

 

Mode No.7 Set the pulse output digit

 

 
If the L/min switch is pressed when Mode No. 6 is lit, the screen will switch to Mode No. 7 and display the following. 
("

" shows the present setting value.) 

 
 
 
 
 
 
 
1.

 

When the L/min switch is pressed for 2 seconds, the setting value will blink, showing ready for setting. 

2.

 

Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is 0, 1, 2, 3, 4, 5, 6, 7, 
or 8.) 

"0" represents non-unit pulse. For setting values other than 0, each time the integrated flow rate display of the 
setting value (i.e. digit) is counted upward, a pulse is outputted.    (e.g., when set to 3, each time the third digit of 
the integrated flow rate display is counted upward, a pulse is outputted.) 

3.

 

When the L/min switch is pressed for 2 seconds, the set value will be lit and setting will be completed. 

Note:

 Set the pulse output digit greater than the most significant digit of pulse constant. 

(If set to a value lower than the most significant digit of pulse constant, pulse output will be indefinite.) 

 
When the L/min switch is pressed in this state, the screen will return to Mode No.1. 
 

 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

6.

 

  

 

Output digit selection 

Mode No.6

 

Mode No.7 

Set the equivalent value. 

- Display the present set value in the light-on state.

Set the EXP value. 

- Display the present set value in the light-on state.

Set the decimal point position of integrated flow rate display and the display unit.

- Display the present set value in the light-on state. 

Set the decimal point position of integrated flow rate (trip) display and the display unit.

- Display the present set value in the light-on state.

Set the decimal point position of instantaneous flow rate display and the display unit.

- Display the present set value in the light-on state.

Set up the upgrade interval of the instantaneous flow rate display

- Display the present set value in the light-on state.

Set the digit of pulse output. 

- Display the present set value in the light-on state.

(Blink) 

Approx. 2 sec

Approx. 2 sec 

Approx. 2 sec

Approx. 2 sec 

Approx. 2 sec

Approx. 2 sec 

Approx. 2 sec 

Approx. 2 sec

Approx. 2 sec

Approx. 2 sec 

Approx. 2 sec 

Approx. 2 sec

Approx. 2 sec 

Approx. 2 sec

(Blink) 

(Blink) 

(Blink) 

(Blink) 

(Blink) 

(Blink) 

.

 



.



 

.

     

 

.

 

 

 

.

 

 

 

.

 

 

 

.

 

 

 

.

 

   

  

.

 

 

 

  

.

 

 

  

 

.

 

 

  

 

.

 

 

   

 

.

     

 

7.

 

 

Содержание NH***-4 Series

Страница 1: ...ndard specifications are as follows NH005 4 10mL P NH010 4 1L P NH020 4 1L P NH05M 4 10mL P NH10M 4 10mL P Common to NH and NK Connect the wires with the meter correctly in accordance with the connect...

Страница 2: ...splay or CL display is blinking the screen will be switched to the integrated flow rate display In switching the integrated flow rate value will not be reset If pressed simultaneously for 2 to 4 secon...

Страница 3: ...display digit on the right will blink 4 Each time the L min switch is pressed the setting value will change in turn Setting value is 1 0 1 2 3 4 or 5 5 When the value is set with the TOTAL TRIP switc...

Страница 4: ...the third digit of the integrated flow rate display is counted upward a pulse is outputted 3 When the L min switch is pressed for 2 seconds the set value will be lit and setting will be completed Not...

Страница 5: ...equivalent value of pulse constant If the SET HOLD switch is pressed when Mode No 1 is lit the screen will switch to Mode No 2 and display the following shows the present setting value 1 When the SET...

Страница 6: ...sec 3 When the SET HOLD switch is pressed for 2 seconds the set value will be lit and setting will be completed When the SET HOLD switch is pressed in this state the screen will return to Mode No 1 N...

Страница 7: ...tructive act such as a crime 3 Caused by corrosion due to use in a corrosive environment 4 Caused by acts of animals such as a dog cat rat or insect 5 Caused by a factor other than our product 6 Which...

Отзывы: