Ex.: When setting the pulse constant using 7.692 mL/P of ND10 per pulse.
7.692mL/P
0.007692L/P
7692
×
10
-6
Mode No.3 (Set the decimal point position of integrated flow rate display)
If the L/min switch is pressed when Mode No. 2 is lit, the screen will switch to Mode No. 3 and display the following.
("
" shows the present setting value.)
1. When the L/min switch is pressed for 2 seconds, the display digit on the left of the setting value will blink,
showing ready for setting.
2. Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is 0 or 1.)
(When set to 0, the value is displayed with the unit of "L", and when set to 1, "m
3
”)
3. When the value is set with the TOTAL/TRIP switch, the display digit on the right will blink.
4. Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is -1, 0, 1, 2, 3, 4, 5, 6,
or 7.)
5. When the value is set with the TOTAL/TRIP switch, the screen will return to the state of "1" above.
6. When the L/min switch is pressed for 2 seconds, the set value will be lit and setting will be completed.
The setting value on the right of the display shows the digit number of the decimal point position. The setting value of
0 means that the least significant digit of the displayed value is the digit of 10
0
L (or 10
0
m
3
), and the setting value of
-1 means that the least significant digit of the displayed value is the digit of 10
1
L (or 10
1
m
3
). The following shows the
relation between setting value and decimal point position.
Mode No.4 (Set the decimal point position of integrated flow rate (trip) display)
If the L/min switch is pressed when Mode No. 3 is lit, the screen will switch to Mode No. 4 and display the following.
("
" shows the present setting value.)
1. When the L/min switch is pressed for 2 seconds, the display digit on the left of the setting value will blink,
showing ready for setting.
2. Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is 0 or 1.)
(When set to 0, the value is displayed with the unit of "L", and when set to 1, "m
3
")
3. When the value is set with the TOTAL/TRIP switch, the display digit on the right will blink.
4. Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is -1, 0, 1, 2, 3, 4, or 5.)
5. When the value is set with the TOTAL/TRIP switch, the screen will return to the state of "1" above.
6. When the L/min switch is pressed for 2 seconds, the set value will be lit and setting will be completed.
The setting value on the right of the display shows the digit number of the decimal point. The setting value of 0
means that the least significant digit of the displayed value is the digit of 10
0
L (or 10
0
m
3
), and the setting value of
-1 means that the least significant digit of the displayed value is the digit of 10
1
L (or 10
1
m
3
).
Mode No.5 (Set the decimal point position of instantaneous flow rate display)
If the L/min switch is pressed when Mode No. 4 is lit, the screen will switch to Mode No. 5 and display the following.
("
" shows the present setting value.)
1. When the L/min switch is pressed for 2 seconds, the display digit on the left of the setting value will blink,
showing ready for setting.
2. Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is 0 or 1.)
(When set to 0, the value is displayed with the unit of "L/min", and when set to 1, "m
3
/min")
3. When the value is set with the TOTAL/TRIP switch, the display digit on the right will blink.
4. Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is -1, 0, 1, 2, or 3.)
5. When the value is set with the TOTAL/TRIP switch, the screen will return to the state of "1" above.
6. When the L/min switch is pressed for 2 seconds, the set value will be lit and setting will be completed.
The setting value on the right of the display shows the digit number of the decimal point position. The setting value
of 0 means that the least significant digit of the displayed value is the digit of 10
0
L (or 10
0
m
3
), and the setting value
of -1 means that the least significant digit of the displayed value is the digit of 10
1
L (or 10
1
m
3
).
.
.
.
.
.
.
.
Mode No.2
Mode No.1
Decimal point position (digit number)
Display unit setting
4.
Display unit setting
Decimal point position (digit number)
Decimal point position (digit number)
Display unit setting
Mode No.5
Mode No.4
Mode No.3
No decimal point
Integrated flow rate display
Not indicated on the LCD.
Equivalent value
EXP value (index)