- 5 - 

- 6-

 

1

7  6  9  2 

2

○.

 

7  6   5   4   3   2  1   0   -1 

Ex.: When setting the pulse constant using 7.692 mL/P of ND10 per pulse. 
 

7.692mL/P 

 0.007692L/P 

 7692 

× 

10

-6

 

 

 
 
 
 
 
 
 

 

Mode No.3 (Set the decimal point position of integrated flow rate display) 

If the L/min switch is pressed when Mode No. 2 is lit, the screen will switch to Mode No. 3 and display the following. 
("

" shows the present setting value.) 

 
 
 
 
1. When the L/min switch is pressed for 2 seconds, the display digit on the left of the setting value will blink, 

showing ready for setting. 

2.  Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is 0 or 1.) 

(When set to 0, the value is displayed with the unit of "L", and when set to 1, "m

3

”) 

3.  When the value is set with the TOTAL/TRIP switch, the display digit on the right will blink. 
4.  Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is -1, 0, 1, 2, 3, 4, 5, 6, 

or 7.) 

5.  When the value is set with the TOTAL/TRIP switch, the screen will return to the state of "1" above. 
6.  When the L/min switch is pressed for 2 seconds, the set value will be lit and setting will be completed. 
 
The setting value on the right of the display shows the digit number of the decimal point position. The setting value of 
0 means that the least significant digit of the displayed value is the digit of 10

0

L  (or 10

0

m

3

), and the setting value of 

-1 means that the least significant digit of the displayed value is the digit of 10

1

L (or 10

1

m

3

). The following shows the 

relation between setting value and decimal point position. 
 
 
 
 
 
 

Mode No.4 (Set the decimal point position of integrated flow rate (trip) display)

 

If the L/min switch is pressed when Mode No. 3 is lit, the screen will switch to Mode No. 4 and display the following. 
("

" shows the present setting value.) 

 
 
 
 
 
1.  When the L/min switch is pressed for 2 seconds, the display digit on the left of the setting value will blink, 

showing ready for setting. 

2.  Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is 0 or 1.) 

(When set to 0, the value is displayed with the unit of "L", and when set to 1, "m

3

") 

3.  When the value is set with the TOTAL/TRIP switch, the display digit on the right will blink. 
4.  Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is -1, 0, 1, 2, 3, 4, or 5.) 
5.  When the value is set with the TOTAL/TRIP switch, the screen will return to the state of "1" above. 
6.  When the L/min switch is pressed for 2 seconds, the set value will be lit and setting will be completed. 

The setting value on the right of the display shows the digit number of the decimal point. The setting value of 0 
means that the least significant digit of the displayed value is the digit of 10

0

L (or 10

0

m

3

), and the setting value of 

-1 means that the least significant digit of the displayed value is the digit of 10

1

L (or 10

1

m

3

). 

 
 

Mode No.5 (Set the decimal point position of instantaneous flow rate display) 

If the L/min switch is pressed when Mode No. 4 is lit, the screen will switch to Mode No. 5 and display the following. 
("

" shows the present setting value.) 

 
 
 
 
 
1. When the L/min switch is pressed for 2 seconds, the display digit on the left of the setting value will blink, 

showing ready for setting. 

2.  Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is 0 or 1.) 

(When set to 0, the value is displayed with the unit of "L/min", and when set to 1, "m

3

/min") 

3.  When the value is set with the TOTAL/TRIP switch, the display digit on the right will blink. 
4.  Each time the L/min switch is pressed, the setting value will change in turn. (Setting value is -1, 0, 1, 2, or 3.) 
5.  When the value is set with the TOTAL/TRIP switch, the screen will return to the state of "1" above. 
6.  When the L/min switch is pressed for 2 seconds, the set value will be lit and setting will be completed. 

The setting value on the right of the display shows the digit number of the decimal point position. The setting value 

of 0 means that the least significant digit of the displayed value is the digit of 10

0

L  (or 10

0

m

3

), and the setting value 

of -1 means that the least significant digit of the displayed value is the digit of 10

1

L (or 10

1

m

3

). 

 

Mode No.2 

Mode No.1

 

Decimal point position (digit number)

Display unit setting 

4. 

 

 

Display unit setting

 

Decimal point position (digit number) 

Decimal point position (digit number) 

Display unit setting 

Mode No.5

Mode No.4

Mode No.3 

No decimal point

Integrated flow rate display 

Not indicated on the LCD. 

Equivalent value 

EXP value (index) 

3.

 

 

 

5.

 

 

 

Содержание NH***-4 Series

Страница 1: ...ndard specifications are as follows NH005 4 10mL P NH010 4 1L P NH020 4 1L P NH05M 4 10mL P NH10M 4 10mL P Common to NH and NK Connect the wires with the meter correctly in accordance with the connect...

Страница 2: ...splay or CL display is blinking the screen will be switched to the integrated flow rate display In switching the integrated flow rate value will not be reset If pressed simultaneously for 2 to 4 secon...

Страница 3: ...display digit on the right will blink 4 Each time the L min switch is pressed the setting value will change in turn Setting value is 1 0 1 2 3 4 or 5 5 When the value is set with the TOTAL TRIP switc...

Страница 4: ...the third digit of the integrated flow rate display is counted upward a pulse is outputted 3 When the L min switch is pressed for 2 seconds the set value will be lit and setting will be completed Not...

Страница 5: ...equivalent value of pulse constant If the SET HOLD switch is pressed when Mode No 1 is lit the screen will switch to Mode No 2 and display the following shows the present setting value 1 When the SET...

Страница 6: ...sec 3 When the SET HOLD switch is pressed for 2 seconds the set value will be lit and setting will be completed When the SET HOLD switch is pressed in this state the screen will return to Mode No 1 N...

Страница 7: ...tructive act such as a crime 3 Caused by corrosion due to use in a corrosive environment 4 Caused by acts of animals such as a dog cat rat or insect 5 Caused by a factor other than our product 6 Which...

Отзывы: