3
© 2022 • Advanced Healthcare Technology Ltd. • BabyPod 2 ®
Section 1.1 - Introduction
This manual contains instructions for initial set-up and inspection of BabyPod 2, proper operator use of the device,
and care and maintenance of BabyPod 2.
Section 1.2 - Warnings
•
Only properly trained personnel who are aware of potential hazards or misuse should use BabyPod 2.
Patients may be harmed by the incorrect use of this device.
•
DO NOT
leave baby unattended at any time while using BabyPod 2.
•
DO NOT
place a baby directly onto the printed side of the TransWarmer Infant Transport Mattress.
TransWarmer’s soft, non-woven surface is designed to be easy on baby’s skin and provide nesting during
transport.
•
DO NOT
lean on BabyPod 2 or place weight on top of BabyPod 2’s lid sections.
•
DO NOT
leave open containers of liquid on or near BabyPod 2. BabyPod 2’s lid section will not prevent fluids
from spilling onto the baby.
•
DO NOT
activate the TransWarmer Infant Transport Mattress in ambient temperatures greater than 28.0
degrees Celsius. Ambient temperatures greater than 28.0 degrees Celsius may cause the mattress to become
too hot and cause injury to the infant.
Section 1.3 - Cautions
•
Always check the security of the Stretcher Fixing Straps prior to transport with BabyPod 2. Strap buckles make
an audible click when fully engaged.
•
DO NOT
modify or make additions to any part of BabyPod 2.
•
Only BabyPod 2 parts and accessories available through customer services and authorised representatives
should be used to replace any components or accessories.
•
Use only non-abrasive cleansers to clean BabyPod 2. Abrasive cleaners may damage the surface of BabyPod 2.
•
Protect BabyPod 2 from the impact of sharp or hard objects that may cause internal damage that is not visible
on the surface of the unit.
•
BabyPod 2 should be inspected before each use for signs of damage to its carbon fibre Outer Shell. Remove
Infant Positioning Straps, Vacuum Mattress and Shock Absorbent Foam and carefully inspect the inner surface
of the carbon fibre shell for signs of stress, including cracks, pockmarks or irregularities. Discontinue use of
BabyPod 2 if signs of stress or damage are present.
•
Chips in the outer shell’s surface due to normal wear and tear generally do not affect the performance of the
BabyPod 2. Damage due to abrasive or rubbing actions may affect the performance and should be carefully
inspected. Discontinue use of BabyPod 2 if signs of abrasive damage are present.
Section 1
General warnings and cautions
Caution
Please read entire instruction manual prior to using BabyPod 2.
Warning
Patients may be harmed by the incorrect use of this device.