11
© 2022 • Advanced Healthcare Technology Ltd. • BabyPod 2 ®
Section 5
Safety inspection of BabyPod 2 cont.
Section 5.3 - Re-assembly of BabyPod 2 cont.
Pass each Infant Positioning Strap through it’s corresponding slot in the Vacuum Mattress. There are
two available slots for each strap to cater for larger or smaller patients.
Insert the “Base” foam section into the bottom of the BabyPod, making sure to keep the 4 Infant
Positioning Straps Accessible. On top of this, place Vacuum Mattress with the vacuum port, where
the feet of the baby will be(recommended), with the valve facing up. The valve should be in the open
position (tubing sections fully engaged with the black “O” ring hidden from view) to allow air to enter
the mattress. Spread the polystyrene bead contents of the mattress evenly by hand.
At this stage, the Vacuum Mattress should be tested according to the instructions in Section 5.4.
Once the Vacuum Mattress has been tested,
place the TransWarmer Infant Transport
Mattress with the printed side up so that
it is ready to be activated. Once activated,
the TransWarmer should be turned so that
the soft, non-woven side faces up.
DO NOT
place a baby directly on the printed side of
the TransWarmer Infant Transport Mattress.