Transducer
Sensor / connector
Measuring range Units
Display
Pulses / print cycle of Mb1
ZA 9909-AK2
0... 65000
S[P]
Alarm value of channel Mb1
beliebig
Alrn
Therm. coeff. M(Mb1)/M(M03)
ZA 9000-FS
W/m
2
K
q/dt
Wet bulb globe temperature
ZA 9000-FS
°C
UbGt
Measured value of Mb1
any
MESS
Cold junction temperature
any
°C
CJ
Number of averaged val. of Mb1
any
n(t)
Volume flow m
3
/h M(Mb1) * Q
any
mh
FLou
TC = Temp. compensation, PC = pressure compensation, Mb1/Mb2 = Reference channels
The use of function channels to output measured values and operands using
the appropriate reference channels Mb1 (normally 1st channel in connector)
and Mb2 (normally M00) is described in the Manual, Section 6.3.4..
Switching off, i.e. deactivating a programmed measuring channel
Function:
RANGE ´
R
´
keys:
,
PROG
CLEAR
The measured value is then no longer displayed, scanned, and output but the
programming is retained intact.
Reactivating the measuring channel:
Function:
RANGE ´
R
´
keys:
,
PROG
PROG
!
If the channel was previously deactivated, then it is now reactivated with all pro-
gramming values intact. However, if the channel is already active, then with
this key combination all its programming values are deleted (i.e. as per se-
lecting the measuring range).
ALMEMO
®
8390-2
41
Sensor programming