Locking mode function
´
LM
´
Select by pressing key:
... Enter as per Sec. 5.4
FUNCTION
1:0 0 0 5
VM
In the display, in front of the locking mode, you should see output function, ele-
ment flags and multiplexer position - assuming these have been programmed
(see Manual, Section 6.10.2/3/4). These can be programmed by pressing and
holding down the
PROG
key.
7.3 Limit values
Two limit values (MAX and MIN) can be programmed per measuring channel.
Exceeding one of these limit values is treated as a fault, i.e. in the same way as
exceeding a measuring range limit or as sensor breakage. In the display the
´ALARM´ arrow appears and the alarm relays are triggered; (see Manual, Secti-
on 6.3.9). Activation of functions ´GH´ and ´GL´, see Section 5.2.
limit value Max function
´
LH
´ and
limit value Min function
´
LL
´
Select by pressing key:
...
FUNCTION
1: 1 2 3.0
GH
Programming:
Enter as per Sec. 5.4
Switch off by keys:
,
PROG
CLEAR
1: - - - -
GH
7.4 Correction values
Sensors can be corrected by means of the correction values ZERO-POINT and
GAIN; (see Manual, Section 6.3.10).
Corrected measured value
= (measured value - zero-point) x gain.
ZERO-POINT CORRECTION function
´
ZC
´
Select by key:
FUNCTION...
Enter as per Sec. 5.4
1: 0 0 3.2
NK
Delete by keys:
PROG
,
CLEAR
1: - - - -
NK
GAIN correction function
´
SC
´
Select by key:
FUNCTION...
Enter as per Sec. 5.4
1:1.5 0 0 0
SK
Once the correction values have been programmed and the measured value
thus duly modified, the ´
CORR
´ arrow appears in the display.
36
ALMEMO
®
8390-2
Sensor programming