background image

1  How to activate or deactivate the Transparent Mode 

 

- 3 – 

 
 

2. 

Press both navigation keys at 
once. 

 

The menu “Transparent Mode” will 
be displayed. 

 

 

3. 

Press the left one of the Navigation 
keys one time. 

 

The Transparent Mode will be 
activated. 

 

The Transparent Mode is active as 
soon as the status display shows 
“ACTIVATED”. 

 

 

 

The ALMEMO

®

 500 app will display the message

 

Transparentmode on!

 

4. 

Connect the data logger via Ethernet or WiFi to a PC (see Instruction 
Manual chapter 7.4.1 Connect PC to the data logger). 

5. 

Open the ALMEMO

® 

Control program. 

6. 

Click on

 Interface.

 

7. 

Click on the tab

 IP Network.

 

8. 

Enter the IP address in the field next to

 IP address or host

 

The IP address is displayed in the status display. 

10. 

In the field next to

 Port

 enter 10001. 

10. 

Click on

 Ok.

 

11. 

Click on

 Search for connected devices.

 

 

When the Transparent Mode is activated, the measuring circuit boards can 
be directly addressed via the ALMEMO

®

 Control program to perform a 

comparison, a multi-point adjustment, or to enable options. 

Deactivate the Transparent Mode 

 

Push one of the Navigation keys. 

Alternatively 

 

Turn the data logger off and then on again. 

Transparent Mode 

Enter? 

Yes 

 

No 

 

Transparent Mode 

ACTIVATED 

192.168.1.204

 

Содержание ALMEMO 500

Страница 1: ...ntationsbild www ahlborn com ALMEMO 500 Supplementary Instruction Manual V7 Technology English V2 0 17 03 2020 Enable transparent mode For comparison multi point adjustment and enabling of options via...

Страница 2: ...connected to the power supply see Instruction Manual chapter 7 1 1 How to connect and check the Power Supply In battery operation no mains adapter connected the Transparent Mode cannot be activated In...

Страница 3: ...ger 5 Open the ALMEMO Control program 6 Click on Interface 7 Click on the tab IP Network 8 Enter the IP address in the field next to IP address or host The IP address is displayed in the status displa...

Страница 4: ...m for the time of activated Transparent Mode the datalogger must be connected to the power supply The tablet must be connected to the data logger see Instruction Manual chapter 6 Initial Operation ste...

Страница 5: ...he Transparent Mode Follow the steps 1 and 2 of the section Activate the Transparent Mode Untick the checkbox next to Transparent Mode by tapping on the checkbox The Transparent Mode will be deactivat...

Страница 6: ...dress or host The IP address is displayed in the status display 6 In the field next to Port enter 10001 7 Click on Ok 8 Click on Search for connected devices Every measuring circuit board will be list...

Страница 7: ...ed 7 8 Click on Search for connected devices Every measuring circuit board will be listed individually in the device list Besides greatest diligence the possibility of incorrect information cannot be...

Страница 8: ...on www ahlborn com under the tab SERVICE on DOWNLOADS Ahlborn Mess und Regelungstechnik GmbH 2020 All rights reserverd Ahlborn Mess und Regelungstechnik GmbH Eichenfeldstra e 1 3 83607 Holzkirchen Ger...