background image

Scaling
The base value and the factor allow for a further scaling of the corrected
measured value of each measuring channel for zero point and slope (gain). The
decimal point position can be set by the exponent. By setting to zero and
entering the nominal value the scaling values can be automatically calculated.

Limit Values and Alarm
Two limit values (1 max and 1 min) can be set for each measuring channel. An
alarm value printout can be performed if a limit value is exceeded and, by
means of relay output modules, alarm contacts are provided that can be
individually allocated to limit values. As a standard, the hysteresis is set to 10
digits, however, it can also be adjusted. Furthermore, limit value exceeding can
also be used to start or stop a data logging.

Sensor Locking
All sensor data stored in the EEPROM of the connector can be protected
against undesired access by means of a graded locking function.

MEASUREMENT
A total of up to 4 measuring channels are available for each transducer, i.e. it is
also possible to evaluate double sensors, individually scaled sensors, or
sensors with function channels. The measuring channels can be successively
selected forwards or backwards via keyboard. The selected measuring point
can be scanned with a conversion rate of 2.5 or 10 measurements/second. The
measured value is calculated and indicated on the display or, if available,
provided on the analogue output.

Measured Value
A continuous presentation of measuring data from the selected measuring point
is provided and also includes automatic zero point correction and optional
correction of the measured value or new scaling.
A sensor breakage condition is, with most sensors, automatically detected
(exception: connectors with shunts, dividers or additional electronics).

Measuring Functions
Special measuring functions are required for some sensors in order to achieve
an optimal acquisition of measuring data. The cold junction compensation is
available for thermocouples, a temperature compensation for dynamic pressure
and pH and conductivity probes, and an atmospheric air pressure
compensation for humidity sensors, dynamic pressure sensors and O

2

 sensors.

With infrared sensors the parameters zero point and slope correction are used
for background temperature and emissivity factor.

Analogue Output and Scaling
By means of analogue start and analogue end the indicated measured value
can be scaled so that the resulting measuring range covers the full analogue
output range (2V, 10V or 20mA).

   6

ALMEMO

 2290-4  

   Functions

Содержание ALMEMO 2290-4 V5

Страница 1: ...Operating Instructions Multifunctional Measuring Instruments ALMEMO 2290 4 with Option Data Logger V1 3 04 12 2003 ALMEMO 2290 4 1...

Страница 2: ...ch On Off Reinitialisation 14 4 CONNECTION OF THE TRANSDUCERS 15 4 1 Transducers 15 4 2 Measuring Inputs and Additional Channels 15 5 DISPLAY AND KEYPAD 16 5 1 Display 16 5 2 Function Selection and Ac...

Страница 3: ...mpensation 36 7 2 Single Measuring Point Scan 36 7 3 Cyclic Measuring Point Scan 37 7 3 1 Print Cycle Continuous Measuring Point Scan 37 7 3 2 Time and Date 38 7 4 Data Memory Option S Data Logger 38...

Страница 4: ...ystem manual section 5 3 All functions and their control via the interface manual section 6 A complete interface command list with all print outputs manual section 7 These operating instructions only...

Страница 5: ...onnected to any ALMEMO measuring instrument and are interchangeable without requiring any settings Function Channels Maximum minimum average values and differences of certain measuring junctions can b...

Страница 6: ...ed forwards or backwards via keyboard The selected measuring point can be scanned with a conversion rate of 2 5 or 10 measurements second The measured value is calculated and indicated on the display...

Страница 7: ...asuring point scan with a time based process flow control is required to digitally register measuring data of all connected sensors For this purpose a real time clock the print cycle and the measuring...

Страница 8: ...ger up to four external output relays and one analogue output Output All measuring and programming data is accessible by means of the LCD display RS232 RS422 RS485 and a Centronics interface are avail...

Страница 9: ...Socket Mains Adapter ZB 2290 NA 12V 200mA Cable electr isol ZB 2290 UK 10 30V 5 LCD Display 6 Function Keys 7 Battery Box back of unit Alkaline mangan battery 9V 6F22 Space or spare battery 5 LCD Disp...

Страница 10: ...display mode DM range R averaging mode AM cross sectional area CS diameter normalized DN atmospheric pressure mb temperature compensation TC locking mode LM limit value Max Hi LH limit value Min Lo L...

Страница 11: ...cycle see 7 3 1 use key M to select the output format in function BAUDRATE BR s 9 2 use key START STOP to start and stop the cyclic measuring point scan s 7 3 6 For storing the measured values only wi...

Страница 12: ...tteries The battery box 7 is located at the underside of the instrument 1 Press the area that is marked with the arrow and at the same time pull as marked by the arrow as illustrated left 2 Use the co...

Страница 13: ...only partly discharged the full capacity cannot be reached by a normal recharging Therefore use the instrument until the rechargeable battery is almost completely discharged and then completely re cha...

Страница 14: ...ed within that period no data will be lost Otherwise a reinitialisation must be performed Reinitialisation If the battery supply has been interrupted or if the device is not functioning properly due t...

Страница 15: ...a disconnection caused by pulling the cable Both levers must be pressed on the sides for disconnecting the connector 4 2 Measuring Inputs and Additional Channels The measuring instrument ALMEMO 2290 4...

Страница 16: ...ed in socket M0 of the device 5 DISPLAY AND KEYPAD 5 1 Display The display 5 of the instrument ALMEMO 2290 4 consists of an LCD module with six and a half 7 segment digits two 16 segment digits and a...

Страница 17: ...bol is illuminated BAT lower than 6 V 1 L o b A t Sensor that are not connected deactivated measuring points cleared programming values Sensor correction or scaling arrow CORR illuminated Measuring po...

Страница 18: ...s are selected by repeatedly operating the keys F1 or F2 If the key is pressed for longer than one second it is possible to switch back to the previous function The functions can be identified by a 2...

Страница 19: ...r connector programming 8 Device programming 9 All functions After a reinitialisation key CLEAR must be pressed when switching on the function group 1 is selected The selection of another function gro...

Страница 20: ...e Flow 1 0834 V Limit Value Max Hi 1 123 0 LH Time 12 34 56 TM Limit Value Min Lo 1 010 0 LL Date 01 12 99 DA Zero Point Correction 1 ZC Start Time 10 00 00 ST Ambient Temperature 1 250 0 AT Start Dat...

Страница 21: ...function keys F1 or F2 or The programming is started by long pressing the key The first programmable digit PROGR flashes and can be altered The digit can be increased using the key After exceeding the...

Страница 22: ...cted measuring channel i e the measurement is not being interrupted Increase the input channel by programmed channels only M Decrease the input channel by press and hold approx 1 sec M 1s 6 2 Selectin...

Страница 23: ...tc type N ZA 9000 FS 30 00 125 00 C Ntc Millivolt 1 ZA 9000 FS 26 000 26 000 mV U 26 Millivolt ZA 9000 FS 10 000 55 000 mV U 55 Millivolt 2 ZA 9000 FS 260 00 260 00 mV U260 Volt ZA 9000 FS 2 6000 2 60...

Страница 24: ...0 00 125 00 C P Ht Rel humidity psychr w PC ZA 9000 FS 0 0 100 0 H P RH Mixture ratio w PC ZA 9000 FS 0 0 500 0 g kg P AH Dew point temperatur w PC ZA 9000 FS 25 0 100 0 C P dt Partial vapour pressure...

Страница 25: ...tters the characters and spaces _ are available The dimension is indicated by two 16 segment characters that are indicated following the measuring and programming values The change of the dimension ca...

Страница 26: ...ed Meas Value Meas Value ZERO POINT x SLOPE Function ZERO POINT CORRECTION ZC Selection with key F2 F2 1 0 0 3 2 ZC Programming Input according to 5 4 Clear with keys long CLEAR PROGR 1 ZC Function SL...

Страница 27: ...display Function EXPONENT EX Selection with key FUNCTION Input see 5 4 1 3 EX Example A force transducer with a 2 0000 V output should indicate 1000 0 N The decimal point must via exponent 3 be shifte...

Страница 28: ...e protected by the locking mode up to an adjustable locking level Before any programming is performed the locking mode must be correspondingly lowered If a dot is indicated following the locking mode...

Страница 29: ...ith the specified conversion rate see man 6 5 4 This is the optimal operating mode for data recording with analogue output 7 1 1 Selecting the Measured Value and Measuring Point After switch on the fu...

Страница 30: ...The peak values are cleared if a total clearing see 7 or change of the range see 6 2 is carried out 7 1 4 Smoothing by Sliding Averaging In case of unstable measured values e g when measuring turbulen...

Страница 31: ...n ALMEMO connector is connected Function Display No averaging Averaging over Measuring Point Scans Continuous averaging over all measuring point scans C o n t Averaging over all measurements of a prin...

Страница 32: ...scan no C in conversion rate 1 Before each measurement the average value must be cleared in the function AVERAGE VALUE by using the keys long CLEAR PROGR 2 In the function MEAS VALUE the single value...

Страница 33: ...ional area is reached Alternatively the average flow velocity can be determined by individual net measurements according to VDI VDE 2640 see 7 1 5 and man 3 5 5 e g 13 24 m s Function AVERAGE VALUE AV...

Страница 34: ...ument is sufficient see above However if the function BA is activated it can be selected with the key F2 and the base value can be cleared with the keys long CLEAR PROGR Function BASE BA with key Clea...

Страница 35: ...ty FY A641 LF 2 77 mS cm FY A641 LF2 147 S cm FY A641 LF3 111 8 mS cm O2 saturation FY A640 O2 101 7 1 8 Atmospheric Pressure Compensation Some measuring variables depend on the environmental atmosphe...

Страница 36: ...cans can be used to acquire indicate and in most cases to document data from the selected meas point and also from other meas points If an interface module is connected to socket A1 the functions PRIN...

Страница 37: ...1 Print Cycle Continuous Measuring Point Scan For cyclic measuring point scans and outputs the print cycle can be set in function PRINT CYCLE PC For activation of functions please refer to section 5...

Страница 38: ...be configured ex factory or subsequently by means of a memory connector with option 2290 4S as a data logger The basic principles for data storage in ALMEMO devices are described in the manual sectio...

Страница 39: ...tivated an S will be indicated at the revision output next to the type of data and the data recording can be immediately started After removing the memory connector the internal memory is also availab...

Страница 40: ...rammed if the functions are activated see 5 2 If no date has been specified the measurement will be performed every day at the same time The current time of day must be already programmed Function STA...

Страница 41: ...ers at least 100 digits e g 0 20mA for 10 0 to 50 0 C To achieve this the analogue output start and the analogue output end of the desired measuring range must be entered within the functions AS and A...

Страница 42: ...and When the desired transmission rate has been selected the programming can be terminated by operating the key The baud rate setting will be stored PROGR in the EEPROM of the interface module and wil...

Страница 43: ...using interface commands or via an additional trigger cable accessory ZA 1000 ET variant 3 see man 6 10 9 This trigger cable is plugged into socket A2 and the outputs are triggered using the external...

Страница 44: ...1 0000 0 1 0000 5 1 MESS 00000 32000 0000 0 1000 0 01 M1 S E2 05 12 0 s man 6 10 1 9 4 Device Address and Networking All ALMEMO instruments can be very easily networked to centrally acquire the measur...

Страница 45: ...st or blockage during operation Remedy Check cabling for inadmissible electrical connection Disconnect external power supply and output modules disconnect suspicious sensors and replace them by hand h...

Страница 46: ...nd possibly control printouts The software AMR Control allows to print the monitor pages including the programming and also to save the terminal operation and to print it out 11 ELECTROMAGNETIC COMPAT...

Страница 47: ...bient air 10 90 rH non condensing Extent of the Delivery Measuring Instrument ALMEMO 2290 4 Operating Instructions ALMEMO 2290 4 ALMEMO Manual incl software AMR Control Product Overview Order No Multi...

Страница 48: ...48 ALMEMO 2290 4 Appendix...

Отзывы: