The sensor is then brought to a defined nominal value (boiling water, known
weight etc.) and the nominal value is entered. For this purpose the key
must twice be pressed long (approx. 1 sec.) until the first digit of the
▲
PROGR
measured value flashes. Then, by entering the nominal value (see 5.4), the
scaling value is calculated, stored as FACTOR and the measured value is
indicated accordingly.
The functions base value ´
BA
´ and factor ´
FA
´ must be activated
and released! (see 6.7)
6.7 Locking the Programming of the Sensor
(s.man. 6.3.12)
The function parameters of each measuring point are protected by the locking
mode up to an adjustable locking level. Before any programming is performed
the locking mode must be correspondingly lowered. If a dot is indicated
following the locking mode on the display then a modification is not possible.
Locking Level
Locked Functions
0
none
1
measuring range + element flags
2
measuring range + zero point and slope correction
3
measuring range + dimension
4
+ zero point and slope correction
5
+ base value, factor, exponent
6
+ analogue output, start and end
7
+ limit values, max and min
Function LOCKING MODE ´
LM
´
Selection with key
F2...
...
Input see 5.4
F2
1:0 0 0 5
LM
If programmed, the output function, element flags and multiplexer setting will be
indicated next to the locking mode (see man. 6.10.2/3/4).
28
ALMEMO
2290-4
Sensor Programming