Function RANGE ´
BE
´
Selection by key
F2...
...
F2
1: N i C r
R
Example :
chann. M1, range NiCr-Ni
Change meas. range:
,
... or
... ,
1s
PROGR
PROGR
Transducer Connector
/
Cable / Sensor
Meas. Range
Dim Display
Pt100-1
ZA 9000-FS
-200.0... +850.0
°C
P104
Pt100-2
ZA 9000-FS
-200.00...+200.00
°C
P204
Ni100
ZA 9000-FS
-60.0... +240.0
°C
N104
NiCr-Ni (K)
ZA 9020-FS
-200.0...+1370.0
°C
NiCr
NiCroSil-NiSil (N)
ZA 9020-FS
-200.0...+1300.0
°C
NiSi
Fe-CuNi (L)
ZA 9000-FS
-200.0... +900.0
°C
FECO
Fe-CuNi (J)
ZA 9000-FS
-200.0...+1000.0
°C
IrCo
Cu-CuNi (U)
ZA 9000-FS
-200.0... +600.0
°C
CUCO
Cu-CuNi (T)
ZA 9000-FS
-200.0... +400.0
°C
CoCo
PtRh10-Pt (S)
ZA 9000-FS
0.0...+1760.0
°C
Pt10
PtRh13-Pt (R)
ZA 9000-FS
0.0...+1760.0
°C
Pt13
PtRh30-PtRh6 (B)
ZA 9000-FS
+400.0...+1800.0
°C
EL18
Au-FeCr
ZA 9000-FS
-270.0... +60.0
°C
AUFE
Ntc type N
ZA 9000-FS
-30.00...+125.00
°C
Ntc
Millivolt 1
ZA 9000-FS
-26.000...+26.000
mV
U 26
Millivolt
ZA 9000-FS
-10.000...+55.000
mV
U 55
Millivolt 2
ZA 9000-FS
-260.00...+260.00
mV
U260
Volt
ZA 9000-FS
-2.6000...+2.6000
V
U2.60
Differential Millivolt 1
ZA 9050-FS
-26.000...+26.000
mV
d 26
Differential Millivolt
ZA 9050-FS
-10.000...+55.000
mV
d 55
Differential Millivolt 2
ZA 9050-FS
-260.00...+260.00
mV
d260
Differential Volt
ZA 9050-FS
-2.6000...+2.6000
V
d2.60
Sensor Voltage
ZA 9000-FS
0.00...20.00
V
UbAt
Milliampere
ZA 9601-FS
-32.000...+32.000
mA
I032
Percent (4-20mA)
ZA 9000-FS
0.00... 100.00
%
P420
Ohm
ZA 9000-FS
0.00... 400.00
Ω
Ohn
Frequency
ZA 9909-AK
0... 25000
Hz
FrEq
Pulses
ZA 9909-AK
0... 65000
PULS
Digital input
ZA 9000-EK2
0.0... 100.0
%
Inp
Digital interface
ZA 9919-AKxx
-65000... +65000
diGi
Infrared 1
ZA 9000-FS
0.0... +200.0
°C
Ir 1
Infrared 2
ZA 9000-FS
0.0... +800.0
°C
Ir 2
Infrared 3
ZA 9000-FS
-30.0... +70.0
°C
Ir 3
ALMEMO
2290-4
23
Sensor Programming
Содержание ALMEMO 2290-4 V5
Страница 48: ...48 ALMEMO 2290 4 Appendix...