Temperature-Monitoring and Error-Diagnosis
Temperature-Monitoring
The monitoring unit displays the measured cooling temperature of the sample gas (black hand) as well as
the defined lower and upper limits (green and red hands).
NOTICE: Temperature display! A dew point within 0°C and +10°C displayed on the control
unit is correct according to the possible working conditions (gas flow-rate, gas inlet
temperature, gas inlet dew point, ambient temperature, etc.). If the dew point goes under 0°C
or over +10°C a malfunction is reported over the potential free alarm contact.
Sample gas cooling temperature (black hand)
Lower limit 0°C (green hand)
Reference value: +3 to +7°C
Upper limit +10°C (red hand)
Error-Diagniosis / Checklist
Fault
Cause
Remedy
Failure of sample gas cooler
Interrupted power supply
Re-establish the power supply
Protective switch activated
Check operating conditions,
reset protective switch manually
Defective refrigerant circuit
Contact AGT Service
Sample gas flow blocked
Gas paths contaminated
Contamination may occure due to dust or
sublimate. Ensure pre-separation
Gas paths frozen
Adhere to operation data
Condensate in gas outlet
Gas inlet/outlet connected incorrectly
Connect gas inlet/outlet correctly
Condensate outlet blocked
Clear condensate outlet,
check installation
Temperature > +10°C
Sample gas cooler overloaded
Check operating conditions,
ensure proper cooling air circulation,
condensor cleaning
Defective refrigerant circuit
Contact AGT Service
Temperature < 0°C
Too low ambient temperature
Adhere to operation data
Defective performance regulation
Contact AGT Service