6.3 Typical Room Layouts
01-999021-00 A0100
Installation Planning Guide
43
Recommended Room Layout—200- and 300-MHz
The room dimensions are about 3 m
×
4.3 m (10 ft
×
14 ft). This arrangement provides
comfortable access for operator and service personnel. A distance of 1.7 m (5 3/4 ft) should
be maintained around the magnet to prevent interference with electronic components. A
distance of 1 m (3 ft) should be maintained around the standard cabinet. Drawn
approximately to scale.
MERCUR
Y
Cabinet
5-gauss
line
{
1/4 in. = 1ft