Chapter 5. Stray Magnetic Fields
40
Installation Planning Guide
01-999021-00 A0100
3.
Magnet area danger signs (
Figure 10
) – Post at each entrance to the magnet area. Be
sure each sign is outside the 5-gauss perimeter.
Stray magnetic fields can reach beyond the published distances when two or more magnetic
fields intersect or when the field extends over large ferromagnetic masses or structures
(steel doors, steel construction beams, etc.). In this case, the customer must measure the
stray field using a gaussmeter to determine how the 5- and 10-gauss fields are altered
(contact a scientific instrumentation supplier for information on acquiring a gaussmeter).
You can request additional signs from Varian by telephoning 1-800-356-4437 in the United
States or by contacting your local Varian office in other countries.
Tools and Equipment
Strong magnetic fields are present that can
make magnetic items suddenly fly towards the
magnet, which could cause personal injury or
serious damage.
Do not take tools, equipment,
or personal items containing steel, iron, or
other magnetic materials closer to the magnet
than this sign.
Dewars
The stray field of the magnet can pull a
magnetic dewar into the magnet body, causing
serious damage. Use only nonmagnetic
stainless steel dewars. Do not use iron or steel
dewars during servicing.
STRONG MAGNETIC FIELD
Pub. No. 87-250302-00 B0694 10-Gauss Warning Sign
10-GAUSS
WARNING
Figure 8. 10-Gauss Warning Sign
Pacemaker, Metallic Implant Hazard
Strong magnetic and rf fields are present that
can cause serious injury or death to persons
with implanted or attached medical devices,
such as pacemakers and prosthetic parts.
Such
persons must not go closer to the magnet than
this sign until safety at a closer distance is
identified by a physician or device
Magnetic Media, ATM/Credit Cards
Strong magnetic fields are present that can
erase magnetic media, disable ATM and credit
cards, and damage some watches. Do not take
such objects closer to the magnet than this sign.
STRONG MAGNETIC FIELD
Pub. No. 87-250303-00 B0694 5-Gauss Warning Sign
5-GAUSS
WARNING
Figure 9. 5-Gauss Warning Sign
STRONG MAGNETIC AND RADIO-FREQUENCY FIELDS ARE PRESENT
Pub. No. 87-250301-00 B0694 Magnet Area Entrance Danger Sign
DANGER
Pacemaker and
Metallic Implant Hazard
Strong magnetic and radio-
frequency fields are present that
could cause serious injury or
death to persons with implanted
or attached medical devices,
such as pacemakers and
prosthetic parts.
Such persons
must not go
closer to the magnet than the
5-GAUSS WARNING signs until
safety at a closer distance is
identified by a physician or
medical device manufacturer.
Tools and Equipment
Strong magnetic fields are
present that could make some
magnetic items suddenly fly
towards the magnet body, which
could cause personal injury or
serious damage.
Do not take tools, equipment,
or personal items containing
steel, iron, or other magnetic
materials closer to the
magnet than the
10-GAUSS WARNING signs.
Magnetic Media and
ATM/Credit Cards
Strong magnetic fields are
present that could erase
magnetic media such as
floppies and tapes, disable ATM
and credit cards, and damage
some watches.
Do not take such objects closer
to the magnet than the
5-GAUSS WARNING signs.
Figure 10. Magnet Area Danger Sign