Agilent U9391C/F Comb Generator Operating and Service Manual
4
Operating Guides
7
Click
Calibrate
button to start the NVNA Cal Wizard.
8
Follow the on screen instruction on the PNA- X to perform
calibration.
9
Vector Calibration
must be performed first. Click
Load
to
recall Cal Set data from a previously performed calibration.
However, the stimulus settings for the Cal set must exactly
match the current stimulus settings.
10
Select Port 1 or Port 3 and follow the instructions to perform
the Amplitude Calibration using power sensor, before doing
the Phase Calibration.
11
To perform
Phase Calibration
, select the phase reference
unit that is identified to use for calibration. Then click
Next
button.
Figure 4-4
NVNA Calibration Wizard