Agilent U9391C/F Comb Generator Operating and Service Manual
33
Quick-Check Procedure
The equipment setup is as illustrated in
1
Click on PNA- X application. Calibrate the network analyzer
with full 2- port calibration using the appropriate electronic
or mechanical calibration kit.
2
Connect Input of comb generator to Port 1 of network
analyzer and Output of comb generator to Port 2 of network
analyzer.
3
Turn ON the biasing of the comb generator by connecting to
power supply, +15V, 0.3A.
4
Get the measurement for Input Return Loss (S11) and Output
Return Loss (S22).
5
Compare measurement results with specification in
Figure 4-1
Quick-check Configuration for S-parameter Test