46
User’s Guide U3042-90002
U3042AE08
Operational Check
NOTE
If you suspect that your 12-Port configuration is not operating properly,
ensure that all front RF jumper interconnect cables are correctly attached.
Equipment Required
The Agilent U3042AE08 requires that the user be familiar with the equipment and
components listed in
Table 16
.
This section provides an equipment list and setup of the analyzer and test set.
Table 15 N5242A Operation Verification Limit for Reflection Tracking
1
1.
Reflection Tracking takes into account Source Loss, Receiver
Loss, Margin, and analyzer’s Mixer Cal.
Frequency
Standard
2
Port 1-12
2. A standard unit is one without Options 001 or 002.
Options
001 or 002
Port 1-12
10 MHz to 4 GHz
−
10.0 dB
0 dB
4 GHz to 6 GHz
−
13.0 dB
−
2.5 dB
6 GHz to 10 GHz
−
16.0 dB
−
5.0 dB
10 GHz to 15 GHz
−
17.0 dB
−
7
.
5 dB
15 GHz to 20 GHz
−
25.0 dB
−
15.0 dB
20 GHz to 26.5 GHz
−
30.0 dB
−
40 dB
Table 16
Equipment List
Description
Qty
N4691A 3.5 mm ECal Module 10 MHz - 26.5 GHz (Option 00F or M0F)
or
N4691B 3.5 mm ECal Module 300 kHz - 26.5 GHz (Option 00F or M0F)
or
Mechanical cal kit 85052B or 85052D
1
N5230C 4-Port Network Analyzer (Option 245 and 551)
or
N5242A Option 400 and 551
1
Set of interconnect cables (analyzer and Test Set), see
“N5230C RF
Interface Cable Connections” on page 15
.
1
This Manual:http://www.manuallib.com/agilent/u3042ae08-test-set-user-guide.html