User’s Guide U3042-90002
19
U3042AE08
System Setup with N5242A
N5242A RF Interface Cable Connections
Figure 17 on page 20
illustrates the setup configuration of the test set and how it should be
configured to the PNA-X. The cables have been supplied with U3021PL3 Option 442.
1. Connect the RF interconnect cables from the PNA-X to the test set in the order listed.
As you are connecting each cable, torque to 8 in-lb. The longer, straight end of each
cable is connected to the test set. Refer to
Table 8
and
Figure 17 on page 20
.
CAUTION
Over torque will cause damage to the test set and may cause connectors to
spin or become loose.
Table 8
N5242A RF Interface Cable Connection
Numeric
Order
RF Cables
From: PNA-X
To: Test Set
1
U3042-20031
Port 1 SOURCE OUT
Port 5 SOURCE IN
2
U3042-20032
Port 1 CPLR THRU
Port 5 CPLR THRU
3
U3042-20029
Port 1 CPLR ARM
Port 5 CPLR ARM
4
U3042-20030
Port 1 RCVR A IN
Port 5 RCVR OUT
5
U3042-20035
Port 3 SOURCE OUT
Port 6 SOURCE IN
6
U3042-20033
Port 3 CPLR ARM
Port 6 CPLR ARM
7
U3042-20034
Port 3 RCVR C IN
Port 6 RCVR OUT
8
U3042-20036
Port 3 CPLR THRU
Port 6 CPLR THRU
9
U3042-20039
Port 4 SOURCE OUT
Port 7 SOURCE IN
10
U3042-20037
Port 4 CPLR ARM
Port 7 CPLR ARM
11
U3042-20038
Port 4 RCVR D IN
Port 7 RCVR OUT
12
U3042-20040
Port 4 CPLR THRU
Port 7 CPLR THRU
13
U3042-20041
Port 2 CPLR ARM
Port 8 CPLR ARM
14
U3042-20044
Port 2 CPLR THRU
Port 8 CPLR THRU
15
U3042-20043
Port 2 SOURCE OUT
Port 8 SOURCE IN
16
U3042-20042
Port 2 RCVR B IN
Port 8 RCVR OUT