8
User’s Guide U3042-90002
U3042AE08
Typical Reflection Tracking
Typical Reflection Tracking
Specifications for the test set are typical. System performance for the analyzer and test set
are only characteristic and intended as non warranted information.
NOTE
Typical specifications are based on 1 to 2 units performance. Refer to
Table 5
,
Table 14
and
Table 15 on page 46
.
Table 5
Typical Reflection Tracking N5230C
Frequency
Standard 700
Option 001
Option 002
300 kHz to 10 MHz
1
1. Generally improves at 3 MHz to
−
6 dBm.
−
80
−
100
−
100
10 MHz to 4 GHz
−
2
+
2
+
1
4 GHz to 6 GHz
−
7
+1
+0
6 GHz to 10.5 GHz
−
8
−
2
−
3
10.5 GHz to 13.5 GHz
−
13
−
3
−
4
13.5 GHz to 15 GHz
−
15
−
3
−
4
15 GHz to 20 GHz
−
20
−
13
−
15
Table 6
Typical Reflection Tracking N5242A
Frequency
Standard 700
Option 001
Option 002
10 MHz to 50 MHz
−
2
+
5
+
4
50 MHz to 500 MHz
−
3
+
5
+
4
500 MHz to 3.2 GHz
−
8
+2
+1
3.2 GHz to 10 GHz
−
12
−
2
−
3
10 GHz to 16 GHz
−
15
−
4
−
5
16 GHz to 20 GHz
−
20
−
13
−
15
20 GHz to 24 GHz
−
24
−21
−
23
24 GHz to 26.5 GHz
−
25
−35
−
37
This Manual:http://www.manuallib.com/agilent/u3042ae08-test-set-user-guide.html