Gas Ballast

The gas ballast port (item 9 on Figure 3 on page 11) is
sealed on the TriScroll 620.

For applications where a lot of water is being pumped,
dry nitrogen at a flow rate of 

5 lpm can be bled into the

gas ballast port. See “Purge Kit” below.

Bearing Purge

The bearing purge port (item 6 on Figure 3 on page 11) is
sealed on the TriScroll 620 and must not be opened or
used.

Purge Kit

A purge kit (Agilent part number PTSPURGEKIT) to
properly purge the gas ballast is available. This kit
contains a flow meter and all necessary valving and
tubing.

CAUTION

The bearing purge port has been 
disabled on the TriScroll 620 model. 
Use of this port could cause internal 
damage to the pump.

Содержание TriScroll 620 Series

Страница 1: ...INSTALLATION AND OPERATION MANUAL Vacuum Products Division DRAFT 10 30 14 TriScrollTM 620 Series Dry Scroll Vacuum Pump Manual No 699904343 Revision G November 2014...

Страница 2: ...TriScroll 620 Series Vacuum Pump Copyright 2014 Agilent Inc DRAFT 10 30 14 TriScroll 620 Dry Scroll Vacuum Pump TriScroll is a trademark of Agilent Inc...

Страница 3: ...earing Purge 16 Purge Kit 16 Troubleshooting 17 Maintenance 18 General Information 18 Related TriScroll Manuals 18 Maintenance and Tooling Kits 18 Factory Service Options 19 Accessories 19 Contacting...

Страница 4: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...

Страница 5: ...ng representatives of Seller which are inconsistent or in conflict with the terms of this warranty shall not be binding upon Seller unless reduced to writing and approved by an officer of Seller Warra...

Страница 6: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...

Страница 7: ...fety protocol Storage When transporting and storing the pump the following environmental requirements should not be exceeded Preparation for Installation The pump is supplied in a special protective p...

Страница 8: ...d of in accordance with specific national and local standards Temperature 5 C to 40 C 41 F to 104 F Relative humidity 0 to 95 non condensing CAUTION As supplied from the factory the pump is configured...

Страница 9: ...Female 3 8 National Pipe Thread NW25 adapter provided Gas ballast Female 1 4 National Pipe Thread 40 Micron sintered filter provided Ambient operating temperature 5 C to 40 C 41 F to 104 F Storage te...

Страница 10: ...ng with Dimensions Figure 2 Pumping Speed Curves 355 13 98 13 50 22 85 Intake NW25 NW25 Exhaust extension included but not installed Exhaust 3 8 NPT port swivels 360 300 11 80 472 18 57 Single Phase 1...

Страница 11: ...Screws 3 13 Motor Electrical Cover Unpacking and Inspection The shipping container is a double carton 1 After opening the outer box remove the foam packing 2 Slit open the inner box 3 Lift the pump wi...

Страница 12: ...the proper capacity The circuit breaker must be suitably located and easily reached The circuit breaker must be marked at the disconnecting device for the equipment with the on off symbol I O and meet...

Страница 13: ...available to strain relieve the electrical supply cable The cable can be mechanically held under the electrical cover in the groove provided The groove is sized for 14 gage cable A 1 2 14 NPSM hole i...

Страница 14: ...s in either direction and realign the two 2 locating pins on the scroll module with the mating frame holes Ensure that the rubber spider is still on the motor coupling and that the coupling teeth are...

Страница 15: ...n it is stopped It will also close when the TriScroll motor thermal overload protection switch shuts down the pump Wiring Using the data in Table 4 verify that the VPI Valve chosen is compatible with...

Страница 16: ...allast port See Purge Kit below Bearing Purge The bearing purge port item 6 on Figure 3 on page 11 is sealed on the TriScroll 620 and must not be opened or used Purge Kit A purge kit Agilent part numb...

Страница 17: ...gth of power supply cable Defective motor Inspect Contact Agilent Poor ultimate pressure System leak Locate and repair leak Water in pump Flush pump with air or dry nitrogen Gas ballast plugged Replac...

Страница 18: ...erform maintenance on TriScroll pumps is provided in kit form A description of each kit and ordering information is provided in Table 7 Table 6 Other Related Manuals Title Applicable TriScroll Model P...

Страница 19: ...ntact Agilent Customer Service at 1 800 882 7426 See the back cover of this manual for a listing of our sales and service offices Visit our web site at http www chem agilent com en US Products Instrum...

Страница 20: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank DRAFT 10 30 14...

Страница 21: ...re product e g inlet screens vent valves x Prior to shipment drain any oils or other liquids purge or flush all gasses and wipe off any excess residue x If ordering an Advance Exchange product p pleas...

Страница 22: ...or been exposed to any toxic or hazardous materials OR HAS pumped or been exposed to the following toxic or hazardous materials If this box is checked the following information must also be filled ou...

Страница 23: ...ther Vacuum leak Other Section 6 ADDITIONAL TERMS Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division Produc...

Страница 24: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...

Страница 25: ......

Страница 26: ...n 54 10040 Leini Torino ITALY Tel 39 011 997 9111 Fax 39 011 997 9350 Toll Free 00 800 234 234 00 Japan Agilent Technologies Japan Ltd Vacuum Products Division 8th Floor Sumitomo Shibaura Building 4 1...

Отзывы: