background image

15

TriScroll 620 Series Vacuum Pump 

DRAFT 10/30/14

Optional Isolation Valve

Scroll pumps return to atmospheric pressure quickly
when shut off, thus the installation of a fast acting,
automatic, normally closed isolation valve is strongly
recommended to prevent pump debris from being
transported back into the vacuum chamber when the
pump is turned off.

The opening of this valve should occur 
simultaneously with or after pump startup.

Valve closing should occur before, 
simultaneously with, but no later than 250 ms 
after pump shut off.

Use an NW25 valve or larger and mount it as close as
possible to the pump inlet. Mounting to the pump inlet is
ideal. (See Figure 5). 

.

Figure 5 Isolation Valve Location

Agilent offers a variety of manual, electromagnetic and
electropneumatic controlled vacuum valves for vacuum
applications. The Agilent Vacuum Pump Isolation (VPI)
Valve is highly recommended for vacuum pump isolation
applications. The VPI valve application and installation
information for use with TriScroll vacuum pumps has
been included below. 

Overview 

The status of the VPI Valve is controlled by the TriScroll
motor internal thermal switch and its electrical supply.
The VPI Valve opens when the pump is running and
closes when it is stopped. It will also close when the
TriScroll motor thermal overload protection switch shuts
down the pump.

Wiring

Using the data in Table 4, verify that the VPI Valve chosen
is compatible with the TriScroll supply voltage and
frequency. Then, locate the proper wiring diagram to use
(Figure 4 on page 13). 

1. Remove the three screws (item 12 in Figure 3 on 

page 11) that are holding the motor electrical cover 
(item 13 in Figure 3).

2. Connect the VPI Valve solenoid wire leads to the 

pump motor as shown in Figure 4.

3. Replace the motor electrical cover and secure it using 

the three screws removed in step 1. Verify that the 
valve is properly grounded before applying 
electrical power.

.

Solenoid operating voltages are lower than the TriScroll vacuum pump operating voltages in order to utilize the TriScroll motor’s inter-

nal thermal switch to actuate the VPI Valve.

Isolation Valve

Pump Inlet

CAUTION

To prevent damage to the VPI Valve, it 
must be installed by a qualified electri-
cian and only as specified below.

Table 4 VPI Valve Installation Data

Motor

TriScroll Vacuum Pump 

Operating Voltage

VPI Valve Part # 

NW25

VPI Valve Part # 

NW40

Solenoid Operating 

Voltage Range 

Use the following 

Wiring Diagram

1 Phase

100-120 V 50/60 Hz

VPI251205060

VPI401205060

90-132 V

Figure 4 low voltage

1 Phase

200-230 V 50/60 Hz

VPI251205060

VPI401205060

90-132 V*

Figure 4 high voltage

NOTE

VPI valve not available on pump model 
X3862-64000.

Содержание TriScroll 620 Series

Страница 1: ...INSTALLATION AND OPERATION MANUAL Vacuum Products Division DRAFT 10 30 14 TriScrollTM 620 Series Dry Scroll Vacuum Pump Manual No 699904343 Revision G November 2014...

Страница 2: ...TriScroll 620 Series Vacuum Pump Copyright 2014 Agilent Inc DRAFT 10 30 14 TriScroll 620 Dry Scroll Vacuum Pump TriScroll is a trademark of Agilent Inc...

Страница 3: ...earing Purge 16 Purge Kit 16 Troubleshooting 17 Maintenance 18 General Information 18 Related TriScroll Manuals 18 Maintenance and Tooling Kits 18 Factory Service Options 19 Accessories 19 Contacting...

Страница 4: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...

Страница 5: ...ng representatives of Seller which are inconsistent or in conflict with the terms of this warranty shall not be binding upon Seller unless reduced to writing and approved by an officer of Seller Warra...

Страница 6: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...

Страница 7: ...fety protocol Storage When transporting and storing the pump the following environmental requirements should not be exceeded Preparation for Installation The pump is supplied in a special protective p...

Страница 8: ...d of in accordance with specific national and local standards Temperature 5 C to 40 C 41 F to 104 F Relative humidity 0 to 95 non condensing CAUTION As supplied from the factory the pump is configured...

Страница 9: ...Female 3 8 National Pipe Thread NW25 adapter provided Gas ballast Female 1 4 National Pipe Thread 40 Micron sintered filter provided Ambient operating temperature 5 C to 40 C 41 F to 104 F Storage te...

Страница 10: ...ng with Dimensions Figure 2 Pumping Speed Curves 355 13 98 13 50 22 85 Intake NW25 NW25 Exhaust extension included but not installed Exhaust 3 8 NPT port swivels 360 300 11 80 472 18 57 Single Phase 1...

Страница 11: ...Screws 3 13 Motor Electrical Cover Unpacking and Inspection The shipping container is a double carton 1 After opening the outer box remove the foam packing 2 Slit open the inner box 3 Lift the pump wi...

Страница 12: ...the proper capacity The circuit breaker must be suitably located and easily reached The circuit breaker must be marked at the disconnecting device for the equipment with the on off symbol I O and meet...

Страница 13: ...available to strain relieve the electrical supply cable The cable can be mechanically held under the electrical cover in the groove provided The groove is sized for 14 gage cable A 1 2 14 NPSM hole i...

Страница 14: ...s in either direction and realign the two 2 locating pins on the scroll module with the mating frame holes Ensure that the rubber spider is still on the motor coupling and that the coupling teeth are...

Страница 15: ...n it is stopped It will also close when the TriScroll motor thermal overload protection switch shuts down the pump Wiring Using the data in Table 4 verify that the VPI Valve chosen is compatible with...

Страница 16: ...allast port See Purge Kit below Bearing Purge The bearing purge port item 6 on Figure 3 on page 11 is sealed on the TriScroll 620 and must not be opened or used Purge Kit A purge kit Agilent part numb...

Страница 17: ...gth of power supply cable Defective motor Inspect Contact Agilent Poor ultimate pressure System leak Locate and repair leak Water in pump Flush pump with air or dry nitrogen Gas ballast plugged Replac...

Страница 18: ...erform maintenance on TriScroll pumps is provided in kit form A description of each kit and ordering information is provided in Table 7 Table 6 Other Related Manuals Title Applicable TriScroll Model P...

Страница 19: ...ntact Agilent Customer Service at 1 800 882 7426 See the back cover of this manual for a listing of our sales and service offices Visit our web site at http www chem agilent com en US Products Instrum...

Страница 20: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank DRAFT 10 30 14...

Страница 21: ...re product e g inlet screens vent valves x Prior to shipment drain any oils or other liquids purge or flush all gasses and wipe off any excess residue x If ordering an Advance Exchange product p pleas...

Страница 22: ...or been exposed to any toxic or hazardous materials OR HAS pumped or been exposed to the following toxic or hazardous materials If this box is checked the following information must also be filled ou...

Страница 23: ...ther Vacuum leak Other Section 6 ADDITIONAL TERMS Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division Produc...

Страница 24: ...TriScroll 620 Series Vacuum Pump This page intentionally left blank...

Страница 25: ......

Страница 26: ...n 54 10040 Leini Torino ITALY Tel 39 011 997 9111 Fax 39 011 997 9350 Toll Free 00 800 234 234 00 Japan Agilent Technologies Japan Ltd Vacuum Products Division 8th Floor Sumitomo Shibaura Building 4 1...

Отзывы: