background image

TriScroll 320 Series Vacuum Pump 

8

Mechanical Connections

Pump Location

Locate the pump on a firm, level surface.

Mounting holes provided in the frame can be used to
provide permanent attachment. The rubber mounts on
the frame can be removed if desired.

Rotation of the Pump Inlet 
with Respect to the Motor Frame

The standard pump configuration is with the inlet fitting
positioned at the top of the pump as shown in Figure 1 on
page 4.
 The pump inlet has two alternate positions:

90 degrees clockwise

90 degrees counterclockwise

To rotate the inlet:

1. Remove three (3) M5 screws (item 1 on Figure 3 on 

page 5) that secure the cowling (item 2 on Figure 3) 
to the scroll module. Remove the cowling.

2. Tilt the pump back so that the rear end of the motor 

touches the floor.

3. Remove the four (4) M6 screws (item 8 on Figure 3) 

holding the frame (item 7 on Figure 3) and scroll 
module together.

4. Axially separate the frame and scroll module.

5. Rotate the module 90 degrees in either direction and 

realign the two (2) locating pins on the scroll module 
with the mating frame holes. Ensure that the rubber 
spider is still on the motor coupling and that the 
coupling teeth are properly aligned.

6. Rejoin the scroll module with the frame and install 

and tighten the four (4) M6 screws removed in step 3.

7. Reinstall the cowling using the three (3) M5 screws 

removed in step 1.

Pump Inlet

Use NW25, or larger, clean vacuum hardware with as
short a length as practical between the pump and the
vacuum chamber.

Use a bellows to provide both vibration isolation and
strain relief between the pump and the vacuum chamber.

Pump Exhaust

A female 1/4" National Pipe Thread exhaust fitting is
located underneath the scroll module. This fitting swivels
360 degrees. Additionally, an NW16 male adapter with
1/4" National Pipe Thread is provided. 

To avoid overheating the pump, do not restrict the
exhaust flow with long lengths of small diameter tubing.
Use as short as practical lengths of NW16 diameter, or
larger, hardware.

Optional Isolation Valve

Scroll pumps return to atmospheric pressure quickly
when shut off, thus the installation of a fast acting,
automatic, normally closed isolation valve is strongly
recommended to prevent pump debris from being
transported back into the vacuum chamber when the
pump is turned off.

The opening of this valve should occur 
simultaneously with or after pump start up.

Valve closing should occur before, 
simultaneously with, but no later than 250 ms 
after pump shut off.

Use an NW25 valve or larger and mount it as close as
possible to the pump inlet. Mounting to the pump inlet is
ideal. (See Figure 5). 

.

Figure 5 Isolation Valve Location

Agilent offers a variety of manual, electromagnetic and
electropneumatic controlled vacuum valves for vacuum
applications. The Agilent Vacuum Pump Isolation (VPI)
Valve is highly recommended for vacuum pump isolation
applications. The VPI valve application and installation
information for use with TriScroll vacuum pumps has
been included below. 

CAUTION

The pump is a heavy object. To avoid 
muscle strain or back injury, use lifting 
aides and proper lifting techniques 
when removing or replacing.

Содержание TriScroll 320 Series

Страница 1: ...INSTALLATION AND OPERATION MANUAL Vacuum Products Division TriScrollTM 320 Series Dry Scroll Vacuum Pump Manual No 699904342 Revision F October 2014...

Страница 2: ...Copyright 2014 Agilent Inc TriScroll 320 Dry Scroll Vacuum Pump TriScroll is a trademark of Agilent Inc...

Страница 3: ...earing Purge 9 Purge Kit 9 Troubleshooting 10 Maintenance 11 General Information 11 Related TriScroll Manuals 11 Maintenance and Tooling Kits 11 Factory Service Options 12 Accessories 12 Contacting Ag...

Страница 4: ...TriScroll 320 Series Vacuum Pump This page intentionally left blank...

Страница 5: ...ocol Storage When transporting and storing the pump the following environmental requirements should not be exceeded Preparation for Installation The pump is supplied in a special protective packing If...

Страница 6: ...ance with specific national and local standards Temperature 5 C to 40 C 41 F to 104 F Relative humidity 0 to 95 non condensing CAUTION As supplied from the factory the pump is configured for low volta...

Страница 7: ...n Female 1 4 National Pipe Thread NW16 adapter provided Gas ballast Female 1 4 National Pipe Thread 40 Micron sintered filter provided Ambient operating temperature 5 C to 40 C 41 F to 104 F Storage t...

Страница 8: ...TriScroll 320 Series Vacuum Pump 4 Figure 1 Interface Drawing with Dimensions Figure 2 Pumping Speed Curves...

Страница 9: ...13 Motor Electrical Cover Unpacking and Inspection The shipping container is a double carton 1 After opening the outer box remove the foam packing 2 Slit open the inner box 3 Lift the pump with the p...

Страница 10: ...er capacity The circuit breaker must be suitably located and easily reached The circuit breaker must be marked at the disconnecting device for the equipment with the on off symbol I O and meet the rel...

Страница 11: ...ct the motor to match your supply voltage Two options are available to strain relieve the electrical supply cable The cable can be mechanically held under the electrical cover in the groove provided T...

Страница 12: ...the vacuum chamber Use a bellows to provide both vibration isolation and strain relief between the pump and the vacuum chamber Pump Exhaust A female 1 4 National Pipe Thread exhaust fitting is located...

Страница 13: ...Scroll motor s inter nal thermal switch to actuate the VPI Valve Gas Ballast The gas ballast port item 9 on Figure 3 on page 5 is sealed on the TriScroll 320 For applications where a lot of water is b...

Страница 14: ...power supply cable Defective motor Inspect Contact Agilent Poor ultimate pressure System leak Locate and repair leak Water in pump Flush pump with air or dry nitrogen Gas ballast plugged Replace breat...

Страница 15: ...maintenance on TriScroll pumps is provided in kit form A description of each kit and ordering information is provided in Table 6 Table 5 Other Related Manuals Title Applicable TriScroll Model Part Num...

Страница 16: ...anual Contacting Agilent In the United States you can contact Agilent Customer Service at 1 800 882 7426 See the back cover of this manual for a listing of our sales and service offices Visit our web...

Страница 17: ...re product e g inlet screens vent valves x Prior to shipment drain any oils or other liquids purge or flush all gasses and wipe off any excess residue x If ordering an Advance Exchange product p pleas...

Страница 18: ...or been exposed to any toxic or hazardous materials OR HAS pumped or been exposed to the following toxic or hazardous materials If this box is checked the following information must also be filled ou...

Страница 19: ...ther Vacuum leak Other Section 6 ADDITIONAL TERMS Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division Produc...

Страница 20: ...TriScroll 320 Series Vacuum Pump This page intentionally left blank...

Страница 21: ......

Страница 22: ...n 54 10040 Leini Torino ITALY Tel 39 011 997 9111 Fax 39 011 997 9350 Toll Free 00 800 234 234 00 Japan Agilent Technologies Japan Ltd Vacuum Products Division 8th Floor Sumitomo Shibaura Building 4 1...

Отзывы: