4

PCR  Programs

 

Gradient Step

48

Agilent SureCycler 8800 Setup and User’s Guide

During the run, the instrument generates a temperature gradient based on 
the values entered into the 

Left Side Temperature Gradient

 and 

Right 

Side Temperature Gradient

 fields. The difference between the minimum 

and maximum temperatures must not exceed 30°C. 

Left Side 

Temperature 

Gradient

Enter the desired temperature for the left side of the thermal block. 

Right Side 

Temperature 

Gradient

Enter the desired temperature for the right side of the thermal block. 

Hold for

Enter the duration of the 

Gradient

 

Step 

(units are 

minutes : seconds

).

Convert to 

Temperature Step 

(optional)

Once the optimal annealing temperature has been determined,

 

the 

Gradient Step

 can be easily converted to a standard 

Temperature

 

Step

 by 

marking this check box and specifying in the drop- down which 
temperature column yielded the best amplification results. The 

Gradient

 

Step

 will then automatically be converted to a standard 

Temperature

 

Step

 

in which the temperature of the thermal block is uniform.

Содержание SureCycler 8800

Страница 1: ...Agilent Technologies SureCycler 8800 Setup and User s Guide Version A March 2011...

Страница 2: ...ate agreement shall control Technology Licenses The hardware and or software described in this document are furnished under a license and may be used or copied only in accor dance with the terms of su...

Страница 3: ...n instructions for setting up the cycler instrument 3 Getting Started This chapter contains instructions to load samples get started with the software set up user accounts and access instruments remot...

Страница 4: ...4 Agilent SureCycler 8800 Setup and User s Guide...

Страница 5: ...stall the thermal block on the instrument base 16 Step 4 Connect the instrument to a power supply 17 Step 5 Connect optional devices 17 Step 6 Turn on the instrument 17 Step 7 Connect the cycler to a...

Страница 6: ...Cycler Settings 32 Hot Top Setting 32 Cycler name 33 Time and date 33 4 PCR Programs 35 Creating New PCR Programs 36 To manually create programs 36 To create a PCR program using the PCR Wizard 39 Tip...

Страница 7: ...Troubleshooting 58 If PCR yield is low when running small reaction volumes with the Standard ramp rate 58 If no actual temperature is displayed during the run 58 If the program does not start 58 If c...

Страница 8: ...8 Agilent SureCycler 8800 Setup and User s Guide Contents...

Страница 9: ...re You Begin Materials Provided 10 Safety Precautions 10 Indicator LED Light 13 Overview 13 SureCycler 8800 Technical Support 14 This chapter contains information for you to read and understand before...

Страница 10: ...urn on the instrument for the first time The device must be connected to a grounded socket Do not touch any switches or outlets with wet hands Turn off the instrument before you disconnect the power c...

Страница 11: ...block inner side of heated lid and reaction vessels These areas quickly attain temperatures of greater than 50 C Keep the heated lid closed until temperatures of 30 C or lower are reached Do not use a...

Страница 12: ...interfere with the normal operation of the USB ports on the instrument Handling precautions are required when working in high static environments Wear a grounded wrist strap and take other antistatic...

Страница 13: ...e most complex thermal cycling techniques including time and temperature increments touchdown PCR and temperature gradients The software interface operated through the color touchscreen features an in...

Страница 14: ...services agilent com World Wide Web www genomics agilent com Telephone USA and Canada 800 227 9770 Europe Japan 0120 477 111 All other countries Go to www genomics agilent com and click on the Contact...

Страница 15: ...e thermal block on the instrument base 16 Step 4 Connect the instrument to a power supply 17 Step 5 Connect optional devices 17 Step 6 Turn on the instrument 17 Step 7 Connect the cycler to a network...

Страница 16: ...pen the shipping containers and unpack the contents Detailed unpacking instructions are provided on the installation poster packed inside the same box as the instrument base Step 3 Install the thermal...

Страница 17: ...ional USB device such as a keyboard mouse or memory stick to the instrument via the USB ports on the front of the instrument 1 Plug the USB cable of the device into either USB port on the instrument S...

Страница 18: ...4 With the buttons firmly pressed lift the thermal block unit up and off the instrument base Use care when lifting the thermal block unit to avoid letting the lid fall closed on your hands Keeping th...

Страница 19: ...User Accounts 24 Access Levels 24 Logging In to User Accounts 24 Managing User Accounts 26 Network Settings and Remote Access 29 To view or edit the cycler s IP address and name 29 To access the cycl...

Страница 20: ...tion plate use a compression mat during cycling to ensure the film stays firmly adhered to the plate See Agilent Plasticware and Consumables on page 21 for ordering information on a compression mat To...

Страница 21: ...The following PCR plates reaction tubes and sealing films are compatible with the SureCycler 8800 Agilent Part Number Description 401333 96 well plate 25 pack 410188 384 well plates 50 pack 410186 Adh...

Страница 22: ...point for all the software controlled instrument operations as well as file and account management Each navigation button on the Home screen menu takes you to a different functional area in the softw...

Страница 23: ...nt SureCycler 8800 Setup and User s Guide 23 Login Opens the Login screen Settings Use this button to adjust cycler settings upgrade software and manage user accounts Table 3 Home Screen Button Descri...

Страница 24: ...In to User Accounts To log in as Guest No password is required to login as Guest By default the Guest account is automatically logged into the system each time the instrument is turned on Table 4 Acce...

Страница 25: ...ect Administrator from the Username drop down list and enter the password The default Administrator password is ADMIN Use the Shift key on the virtual keyboard to toggle between lowercase and uppercas...

Страница 26: ...d for that account If logging in as Guest leave the Password text box blank 4 Click or touch the Login button A confirmation message appears to verify the name of the user logged in to the system Mana...

Страница 27: ...field enter a user name 5 In the Password and Confirm Password fields enter a password for the account 6 Select the access level User or Administrator from the Access Level drop down list Selecting Us...

Страница 28: ...ccount Note that the access level for the Administrator and Guest accounts cannot be edited 5 Press OK to save the changes A confirmation message appears after successful update of the user data To de...

Страница 29: ...and the DHCP server You have the option of obtaining an IP address automatically or assigning one manually If you select the option Obtain an IP Address automatically the fields for network informatio...

Страница 30: ...me of the instrument you want to connect to into the IP Address Cycler field 4 Press Connect You will be returned to the Network Control screen and the network information and status of the connected...

Страница 31: ...access the cycler from a PC 1 From the PC open an internet browser application Microsoft Internet Explorer is recommended 2 In the URL field enter the IP address for the cycler The SureCycler 8800 so...

Страница 32: ...ting screen make a selection for use of the hot top Select Yes to use the hot top when the instrument is running Select No to turn the hot top off 4 Confirm your selection by pressing OK When the hot...

Страница 33: ...ress System Setting 3 Press the Change button located near the top of the screen A new screen will open prompting you to enter a new cycler name 4 Confirm your changes by pressing OK Time and date The...

Страница 34: ...d Time and date 34 Agilent SureCycler 8800 Setup and User s Guide 4 To change the time press the up down arrows to adjust the hour minutes or seconds listed under System Time 5 Confirm your changes by...

Страница 35: ...PCR Wizard 39 Tips for Creating PCR Programs for the 384 well SureCycler 8800 42 Program Steps in the Commands List 43 Using a Preloaded PCR Program 51 Editing PCR Programs 53 Running a PCR Program 5...

Страница 36: ...program automatically The two methods are described in the following sections To manually create programs Programs can be created manually from the programming screen The steps in the Command List ar...

Страница 37: ...e first step of the program from the Command List to the Program Window using one of the following two approaches Press the command then press the forward arrow button shown below Double press the com...

Страница 38: ...e Save Program screen will open b Press the keyboard icon shown below left to access the touchscreen s virtual keyboard and use the keyboard to type a name into the Program Name field Once the name ha...

Страница 39: ...of the PCR Wizard 1 Polymerase Selection Select a DNA polymerase from the list and press Next The polymerase options are Paq5000 hot start or standard version PfuUltra II Herculase II PfuTurbo hot st...

Страница 40: ...s If you entered the base sequences of the primers rather than the Tm s the software calculates the Tm for each primer based on the formula Tm 69 3 0 41 GC 650 N where GC is the percentage of GC conte...

Страница 41: ...riate gradient range based on the melting temperatures of the primers Once the wizard has guided you through these five steps it creates a PCR program based on the provided information and displays it...

Страница 42: ...well block reducing the duration of the denaturation step during cycling may improve product yield for certain enzymes Denaturation times as short as 3 5 seconds may be optimal in some protocols If y...

Страница 43: ...es with hot start technology The extended denaturation step activates the polymerase by either removing a chemical modification agent or by denaturing a polymerase bound antibody The Hot Start command...

Страница 44: ...Step Enter the denaturation temperature in C left field and denaturation time in hours minutes seconds right field Annealing Step Enter the annealing temperature in C left field and annealing time in...

Страница 45: ...before a series of steps that are to be repeated for several cycles Number of Cycles Enter the number of times the series of steps should be cycled Cycle Name Optional If desired enter a user selecte...

Страница 46: ...er the temperature for the step in C Hold for Enter the duration of the step Units are hours minutes seconds Temperature Ramp Select the ramp rate for the instrument The Fast ramp rate is the default...

Страница 47: ...Start Cycle and an End Cycle command Insert this program step to terminate a PCR cycle Put this step after a series of steps that are to be repeated for several cycles Gradient Step A Gradient Step i...

Страница 48: ...ermal block Right Side Temperature Gradient Enter the desired temperature for the right side of the thermal block Hold for Enter the duration of the Gradient Step units are minutes seconds Convert to...

Страница 49: ...aximum starting annealing temperature and the minimum ending annealing temperature Then during the run the software executes the cycling protocol so that the annealing temperature is lowered by an equ...

Страница 50: ...nd Hold Time Enter the elongation temperature in the Elongation Temperature field and the duration of elongation in the adjacent Hold For field to be used during touchdown cycling Cycles Enter the num...

Страница 51: ...ler 8800 PCR Program File Name PCR Enzyme or Kit Denaturation Annealing Elongation Herculase II Fusion 1kb target pcr Herculase II Fusion DNA Polymerase 95 C for 20 seconds 55 C for 20 seconds 72 C fo...

Страница 52: ...ct it Press Run to run the selected program or Open to view the program parameters in the programming screen To edit a preloaded program 1 Press the Programs button on the Home screen 2 In the browser...

Страница 53: ...ler 8800 instrument To edit program step parameters 1 Press the Programs button on the Home screen Depending on your user level Guest or registered user the specific user directory or Guest directory...

Страница 54: ...the new step See Program Steps in the Commands List on page 43 for details on parameters for a particular command To delete a step from the program press the command in the Program Window to select it...

Страница 55: ...eactions in microliters and press OK From the programming screen 1 Press Start 2 In the dialog box enter the volume of the reactions in microliters and press OK To stop or pause a program From the Sta...

Страница 56: ...ncubation 1 From the Home screen press Incubation 2 In the Incubation Temperature field enter the temperature of the incubation 3 Specify the duration of the incubation If you want the instrument to h...

Страница 57: ...opens the Browse Reports screen listing the report files 2 You can scroll through the list using the arrow buttons at the bottom of the screen or sort the reports by name or date Press the Report Name...

Страница 58: ...ng Denaturation times as short as 3 5 seconds may be appropriate Alternatively switch to using capped reaction tubes Increase the reaction volume to 50 l Try switching to the Fast ramp rate If no actu...

Страница 59: ...he hot top is disabled Enable the hot top using the instructions in Hot Top Setting on page 32 If the sealed microplate loses sample volume Check that the microplate sealing system is of good quality...

Страница 60: ...and unlit Check that the power is on at supply and the power cord is connected Check fuses at rear of instrument above power inlet If screen is lit but shows no information or has missing characters R...

Страница 61: ...nd User s Guide Agilent Technologies 5 Maintenance and Updates Maintenance 62 Cleaning 62 Replacing a fuse 62 Software Updates 63 This chapter contains instructions for maintaining the instrument and...

Страница 62: ...ent Do not let liquid enter the instrument Turn off and disconnect the instrument from the power supply before you clean Replacing a fuse WARNING Fuse compartments are located on the rear of the instr...

Страница 63: ...ator on page 25 for instructions 3 From the Home screen press Settings and then press Software Update 4 Insert the USB memory stick into either of the USB ports on the front of the cycler 5 Press Brow...

Страница 64: ...5 Maintenance and Updates Software Updates 64 Agilent SureCycler 8800 Setup and User s Guide...

Страница 65: ......

Страница 66: ...Agilent Technologies Inc 2011 Version A March 2011 G8800 90000 G8800 90000 www agilent com Agilent Technologies In This Book This document describes how to program and use the Agilent SureCycler 8800...

Отзывы: