4
PCR Programs
Gradient Step
48
Agilent SureCycler 8800 Setup and User’s Guide
During the run, the instrument generates a temperature gradient based on
the values entered into the
Left Side Temperature Gradient
and
Right
Side Temperature Gradient
fields. The difference between the minimum
and maximum temperatures must not exceed 30°C.
Left Side
Temperature
Gradient
Enter the desired temperature for the left side of the thermal block.
Right Side
Temperature
Gradient
Enter the desired temperature for the right side of the thermal block.
Hold for
Enter the duration of the
Gradient
Step
(units are
minutes : seconds
).
Convert to
Temperature Step
(optional)
Once the optimal annealing temperature has been determined,
the
Gradient Step
can be easily converted to a standard
Temperature
Step
by
marking this check box and specifying in the drop- down which
temperature column yielded the best amplification results. The
Gradient
Step
will then automatically be converted to a standard
Temperature
Step
in which the temperature of the thermal block is uniform.
Содержание SureCycler 8800
Страница 1: ...Agilent Technologies SureCycler 8800 Setup and User s Guide Version A March 2011...
Страница 4: ...4 Agilent SureCycler 8800 Setup and User s Guide...
Страница 8: ...8 Agilent SureCycler 8800 Setup and User s Guide Contents...
Страница 64: ...5 Maintenance and Updates Software Updates 64 Agilent SureCycler 8800 Setup and User s Guide...
Страница 65: ......