![Agilent Technologies SureCycler 8800 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/agilent-technologies/surecycler-8800/surecycler-8800_setup-and-user-manual_2867794018.webp)
2
Installation
Thermal Block Exchange
18
Agilent SureCycler 8800 Setup and User’s Guide
Thermal Block Exchange
Two interchangeable thermal block units are available with the SureCycler
8800: the 96- well block and the 384- well block.
To remove the thermal block
1
Power off the instrument and disconnect the power supply.
2
Lift the lid off the thermal block by pressing firmly on the gray button
on the top of the instrument.
3
Using both hands, grasp the thermal block unit on the left and right
sides while using your thumbs to press down on the buttons in the
upper left and right corners.
4
With the buttons firmly pressed, lift the thermal block unit up and off
the instrument base.
Use care when lifting the thermal block unit to avoid letting the lid
fall closed on your hands. Keeping the unit tilted slightly away from
you may help prevent the lid from falling closed.
To install the thermal block
1
Using both hands, grasp the thermal block unit on the left and right
sides and place it on top of the instrument base.
2
Firmly press down on both sides of the thermal block unit until it locks
into place. Ensure both the left and right sides of the thermal block
unit are locked in place by pressing down on both sides.
The thermal block unit will only lock into place if it is placed on the
instrument base in the proper orientation.
3
Reconnect the power supply when finished.
Содержание SureCycler 8800
Страница 1: ...Agilent Technologies SureCycler 8800 Setup and User s Guide Version A March 2011...
Страница 4: ...4 Agilent SureCycler 8800 Setup and User s Guide...
Страница 8: ...8 Agilent SureCycler 8800 Setup and User s Guide Contents...
Страница 64: ...5 Maintenance and Updates Software Updates 64 Agilent SureCycler 8800 Setup and User s Guide...
Страница 65: ......