![Agilent Technologies N4000A Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/agilent-technologies/n4000a/n4000a_operating-and-service-manual_2868361048.webp)
38
Chapter 2
ENR File Format
Format Details
Frequency fields
The frequency field (
Freq
) is the frequency at which the ENR amplitude
was measured.
The frequency unit field (
Funit
) is optional. The default unit is Hz.
Valid units are Hz, kHz, MHz, GHz, THz. Units are not case sensitive.
ENR fields
The ENR amplitude (
ENR
) is the measured ENR at the specified
frequency.
The ENR unit field (
Eunit
) is optional. The default unit is dB. Currently,
the only allowed unit is dB. Note that units K, C, F (temperature) are
reserved for possible future use, but are not supported by the NFA at this
time.
The uncertainty field for the ENR amplitude (
Euncert
) is optional.
However, this field must be present if reflection coefficient data is
supplied.
Reflection
coefficient data
The reflection coefficient data is optional and is formatted as four fields
(
on_mag, on_phase, off_mag, off_phase
):
• Reflection magnitude with noise source on.
• Reflection angle (in degrees) with noise source on.
• Reflection magnitude with noise source off.
• Reflection angle (in degrees) with noise source off.
If any reflection coefficient data is supplied, then all four fields must be
present.
NOTE
The file format requires the ENR uncertainty field to be present before
any reflection data.
The reflection coefficient uncertainty (
on_mag_uncert
) is an optional
parameter. Either one or four reflection uncertainties are allowed. If one
value is supplied, it is assumed to apply to all reflection parameters.
Alternatively, uncertainties can be supplied for all four reflection
parameters independently. Reflection coefficient data must be present if
this field is supplied.