Operator’s Check for Switching Noise Source Off and On
The following procedure checks that the SNS can switch the noise source
off and on. This ensures the excess noise switches.
Step 1. Remove any cables from the NFA input and SNS drive.
Step 2. Press the
Preset
key.
Wait until the preset routine has completed before proceeding.
NOTE
The preset condition used is the NFA default of
Preset (Factory)
.
Step 3. Connect the SNS to the NFA Port using the multi-pin 11730 cable, as
shown in Figure 1-12.
Figure 1-12
SNS connected to NFA graphic
Step 4. Press the
Frequency/Points
key.
Step 5. Press the
Frequency Mode
menu key.
Step 6. Press the
Fixed
menu key.