18
Chapter 1
General Information
Operation
Operator’s Check for T
cold
Communication
The following procedure checks that the SNS can transfer the T
cold
data
to the NFA.
Step 1. Remove any cables from the NFA input and SNS drive.
Step 2. Press the
Preset
key.
Wait until the preset routine is complete before proceeding.
NOTE
The preset condition used is the NFA default of
Preset (Factory)
.
Step 3. Press the
ENR
key.
Step 4. Press the
Tcold
menu key.
Step 5. Connect the SNS to the NFA Port using the multi-pin 11730 cable, as
shown in Figure 1-8
Figure 1-8
Operator’s Check Test Setup for T
cold
Communication
Step 6. Press the
SNS Tcold
menu key and set it to
SNS Tcold (Off)
.