IDP-15 Dry Scroll Vacuum Pump
47
DRAFT
12
/21/17
To test the pump:
1. Run the pump for about 5 seconds. Verify that the front fan is running.
If you hear loud noises or observe labored operation, this indicates that the tip seal or
main O-ring are possibly out of place.
2. Disassemble and repair as necessary.
The pump is now ready to return to service.
NOTE
Newly installed tip seals may require several hours of run time
to seat properly and enable the pump to meet speed and
base pressure specifications.