Agilent 7697A Headspace Sampler
G4565A Cooling Plate Accessory Installation Guide
11
To power on the headspace sampler
1
Power on the headspace sampler (
2
Press
[Status] > [Clear]
on the headspace keypad to display the firmware
revision. If the firmware revision is lower than A.01.05.1, you must
update the headspace firmware. See
for details.
To update the headspace firmware
The 7697A Headspace Sampler firmware version must be A.01.05.1 or
greater to control the G4565A Cooling Plate Accessory.
Use Agilent Instrument Utilities software, version B.01.05 SP1 or later to
update the headspace firmware.
1
Load Agilent Instrument Utilities.
2
Select
Firmware Update
from the explorer pane.
3
Select the
Agilent 7697A Headspace Sampler
from the
Instruments
list
and click
Connect
.
If the Agilent 7697A Headspace Sampler is not available, click
Add New
Instrument
to open
Configure Instruments
. Add the new instrument, then
return to firmware update.
Figure 14
Power on the headspace sampler