4
G4565A Cooling Plate Accessory Installation Guide
Agilent 7697A Headspace Sampler
Installation
Read the Agilent 7697A Headspace Sampler Safety Manual before installing
the cooling plate accessory, and whenever needed. The safety manual
describes precautions to follow when performing maintenance and using
the headspace sampler, as well as important regulatory information.
Confirm that all parts have been received. Refer to
and
for information.
To prepare the GC and 7697A 111-Vial Headspace Sampler
1
Set your GC oven, GC inlet, headspace oven, and transfer line to
ambient temperatures and wait for them to cool.
2
Once the GC oven, headspace oven, and transfer line have cooled to
ambient temperature, turn off the headspace and unplug the power
cord.
3
Turn off all gas flows at their sources.
4
Remove all sample vials and vial racks from the headspace tray area.
C A U T I O N
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.