G3388B Leak Detector
17
Replacing the filters
The detector includes filters on the sample probe and
reference port to protect against particulate
contamination. A decrease in sensitivity can indicate a
clogged filter. Replace the filters as described below.
•
Replacement sample probe filter: G3388-80001
•
Replacement reference port filter: G3388-80002
To replace the sample probe filter
1
Turn off the detector.
2
Unscrew the probe halves and disassemble the
probe.
3
Use long-nosed pliers to remove the old filter.
4
Install the filter into the probe base as shown below.
Sampling probe
Filter
Probe base
Probe tip