Detecting leaks
To use the detector:
1
Press and hold [
FUNC/POWER
] to turn on the detec-
tor. You will hear the detector begin to sample air.
Wait until the warmup completes.
2
If desired, set the sensitivity range or other settings.
See
“Setting options”
on page 13.
3
Check the stability at the zero point (while sampling
clean air). When stable, the leak indicator in the LCD
display will alternate between
■
and
□
.
•
If the detector indicates a leak (more than one
bar), press [
ZERO/ENTER
] to reset the zero bal-
ance.
•
Temperature or humidity fluctuations can impact
the zero point.
4
Place the sample probe near the gas connection to
check for a leak. Do not press the probe tip against
the fittings, tubing, and so forth.
•
One bar (or LED) may not indicate a leak, espe-
cially in high sensitivity mode.
•
Two or more bars indicates a leak.
•
An audible alarm sounds when 3 or more bars
display (with Buzzer set to On).
Turn off the detector when not in use. Press
FUNC/POWER
and hold for 3 seconds.
The detector can be set to turn off automatically after
10 minutes. To enable Auto Power Off, see
“Setting
options”
on page 13.