1200 Series Handheld Control Module Users’s Guide
69
Using the Pump
3
Tracking Pump History / Pump EMF Limits
Tracking Pump History / Pump EMF Limits
1
Select the
Records
button [
F4
] in the
System
screen.
2
Select Pump from the list.
3
Press
Enter
.
You can now check the amount of solvent the pump has delivered (liquimeter)
and the wear count.
The liquimeter displays the total volume of solvent that the pump has
delivered since it was last reset. You can use the liquimeter limit to set up a
preventive maintenance schedule for the pump. For example, make a note of
the total number of liters pumped when you change the piston seals. This
volume is a benchmark, reflecting the lifetime of the seals for your solvents
and application. When the limit is exceeded an early maintenance feedback
message will appear indicating it is time to change the seals. Changing the
seals in advance will prevent you having to repeat analyses, due to leaking
seals.
The wear count displays a calculated value indicating the wear of the pump
seals. You can also use the seal wear limit to set up a preventive maintenance
schedule for changing the pump seals when the limit is exceeded.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......