1200 Series Handheld Control Module Users’s Guide
41
Working with the Control Module
2
Working with Methods
Working with Methods
A method contains a complete set of injection, separation and detection
parameters, including the timetable and injector program. Vial range
information is not part of the method.
There are two types of methods:
•
The module method. The method parameters are stored in the individual
LC modules and not in the control module. A method that is stored in the
individual LC modules can be loaded, modified, saved and run from the
control module.
•
The PC card method. The method parameters are stored on a PC card. A
method that is stored on the PC card can be loaded to the LC modules or
transferred to another LC system. Methods cannot be run directly from the
PC card. The method must first be loaded from the PC card before it can be
run. When the PC card method is loaded it becomes a module method.
Unless stated otherwise, the following sections refer to module methods.
Loading a Method
A method can be loaded using the Method button [F3] in the Analysis or
Samples screen:
1
Enter the
Analysis
or
Samples
screen.
2
Select
Method
; the current parameters are displayed.
3
Press the
Module
button [
F1
].
4
Select a method from the list.
5
Press
Enter
.
You can also load a method directly from the
Analysis
screen by using the
selection keys and changing the method in the method list.
The
Method/Module
screen lists all methods that are stored in the modules. For
each method there is a date when the method was last changed and a short
user description. When a method is loaded it becomes the current method.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......