![Agilent Technologies ESG series Скачать руководство пользователя страница 262](http://html.mh-extra.com/html/agilent-technologies/esg-series/esg-series_user-and-programming-manual_2867455262.webp)
4-18
Real-Time I/Q Baseband Generator User’s and Programming Guide
Programming Commands and Examples
ESG Family Signal Generators
Programming Commands
Options UN8 and 202
P32
Thirty-two 1’s and thirty-two 0’s
P64
Sixty-four 1’s and sixty-four 0’s
*RST value: PN9
GSM - Normal Timeslot Encryption Data Field
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:NORMal:ENCRyption
PN9|PN15|FIX4|"<file name>"|EXT|P4|P8|P16|P32|P64|MPN9|MPN15
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:NORMal:ENCRyption?
This command enables you to create and configure an encrypted data field for a normal
timeslot. The choices are timeslots 0 through 7 and PN9 (pseudorandom bit pattern), PN15
(pseudorandom bit pattern), FIX4 (4-bit repeating sequence data pattern), "<file name>" (a
file in the catalog of binary files), Ext (external input data pattern), P4 (four 1’s and four
0’s), P8 (eight 1’s and eight 0’s), P16 (sixteen 1’s and sixteen 0’s), P32 (thirty-two 1’s and
thirty-two 0’s), P64 (sixty-four 1’s and sixty-four 0’s), MPN9 (inserts a PN9 sequence into a
multiframe slot), or MPN15 (inserts a PN15 sequence into a multiframe slot).
*RST value: PN9
GSM - CS1 Normal Timeslot Encryption Data Field
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:NORMal:ENCRyption:CS1:
DATA PN9|PN15
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:NORMal:ENCRyption:CS1:DATA?
This command enables you to create and configure a block type 1 encrypted data field for a
normal timeslot.
*RST value: PN9
GSM - Downlink Normal Timeslot Encryption Data Field
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:NORMal:ENCRyption:DLINk:MCS1:
DATA PN9|PN15
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:NORMal:ENCRyption:DLINk:MCS1:
DATA?
This command enables you to create and configure a downlink block type 5 encrypted data
field for a normal timeslot.
*RST value: PN9
GSM - TCH/FS Normal Timeslot Encryption Data Field
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:NORMal:ENCRyption:TCH:FS:
DATA PN9|PN15
[:SOURce]:RADio:GSM:SLOT0|[1]|2|3|4|5|6|7:NORMal:ENCRyption:TCH:FS:DATA?
This command enables you to create and configure a full rate speech encrypted data field
for a normal timeslot.
*RST value: PN9