Installation Note E4411-90051
Remove Rear Frame and Outer Cover
1. Disconnect the ac power cord from the spectrum analyzer.
2. Refer to Figure 1. Carefully put the analyzer on the work surface with the front frame, (2) and (3),
facing down.
3. Position the handle (4) as shown in the figure.
4. Remove the four screws (5) that hold the rear frame (6) and outer cover (7) in place.
5. Lift the instrument rear frame (6) and outer cover (7) up and off of the instrument and set them
aside.
Figure 1.
Remove Rear Frame and Outer Cover