Getting Started
17
Chapter 1
Disabling the
On/Stdby Switch
The front panel On/Stdby switch is disabled by removing surface mount 0
resistor located on the front monitor board (see Figure 1-6). The resistor is
labeled:
REM PWR JUMPER
.
Caution
After removing the 0
resistor, heat damage may prevent the
resistor from being re-installed to re-enable the On/Stdby
switch.
To access the basic monitor board and resistor:
1. Turn off the mainframe and remove the power cord.
2. Remove the rear panel. Refer to “Removing the Rear Panel from the
Mainframe” on page 45 for instructions.
3. Remove the mainframe cover by removing the 10 (five on each side)
m3x6 flat head torx screws.
Figure 1-6. Disabling the On/Stdby Switch
Location of 0