Installation Note E8364-90012
About Installing the Upgrade Kit
Description of Option 014
An Option 014 analyzer can be configured to measure high-power devices and devices for high
dynamic range.
For a high-power measurement, external amplifiers and high power attenuators or isolators
can be added to complete the test setup. In this configuration, test port output power up to
1 Watt (+30 dBm) can be applied to the device under test (DUT). Additionally, there is an
external reference input that allows the external amplifier’s frequency response and drift to
be ratioed out.
For high dynamic range measurements, front panel jumpers are moved to reverse the signal
path through one of the couplers. This allows for a 15 dB improvement in transmitted signal
sensitivity in one direction only. These jumpers are installed on both ports allowing the user
to choose a high dynamic range measurement in either the forward or reverse direction.
Products affected . . . . . . . . . . . . . . . . . . . . . . . . . . E8363A and E8364A; all options
Installation to be performed by . . . . . . . . . . . . . . Agilent service center or personnel
qualified by Agilent
Estimated installation time . . . . . . . . . . . . . . . . . 2.0 hours
Estimated adjustment time . . . . . . . . . . . . . . . . . 0.5 hours
Estimated full instrument calibration time . . . . 4.5 hours