Installation Note E8364-90012
Analyzers with Option UNL
Refer to
for the following procedure. The new parts referenced in this procedure are
listed in
If you are installing Option 014 on an analyzer that has Option UNL installed, install the
following cables in the order listed:
• W70 E8364-20049 REFERENCE 1 RCVR R1 IN to A28 channel R1 mixer
• W71 E8364-20050 REFERENCE 2 RCVR R2 IN to A29 channel R2 mixer
• W68 E8364-20044 A26 test port 1 coupler to PORT 2 CPLR ARM
• W72 E8364-20046 PORT 2 RCVR B IN to A30 channel B mixer
• W82 E8364-20054 A37 step attenuator to PORT 2 SOURCE OUT
• W84 E8364-20040 PORT 2 CPLR THRU to A39 bias tee
• W66 E8364-20048 A24 detector to REFERENCE 2 SOURCE OUT
• W67 E8364-20043 A25 test port 1 coupler to PORT 1 CPLR ARM
• W69 E8364-20045 PORT 1 RCVR A IN to A27 channel A mixer
• W81 E8364-20053 A36 step attenuator to PORT 1 SOURCE OUT
• W83 E8364-20039 PORT 1 CPLR THRU to A38 bias tee
• W65 E8364-20047 A23 detector to REFERENCE 1 SOURCE OUT
Figure 8
Option 014 Cable Installation, Analyzers with Option UNL