Service Guide E8364-90038
7- 67
PNA Series Microwave Network Analyzers
Repair and Replacement Procedures
E8362C, E8363C, E8364C
Removing and Replacing the A41 Hard Disk Drive (HDD)
Initialize the Windows Operating System
1. Connect a mouse and keyboard to the analyzer.
2. Turn on power to the analyzer.
3.
The operating system initialization process will run automatically. Part way through the process, you
will be prompted to accept the Windows licensing agreement. The entire process takes approximately
six minutes, during which the analyzer will reboot up to three times.
4.
After the PNA application screen is displayed, there may be a display to remind you to upgrade the
firmware. If so, click
OK
to continue.
Set Up the Agilent Administrator Password
After an HDD replacement, it is necessary to set up the Agilent administrator password for the analyzer.
Go to the following Internet site, enter your PNA serial number, and get a six-digit numerical password:
http://na.tm.agilent.com/pna/hdd/index.html#agpswd
.
The network analyzer must be powered up and operating for this procedure.
1.
On the
Windows Desktop
, right-click on
My Computer
, and then click
Manage
.
2.
In the
Computer Management
window, navigate to
System Tools
,
Local Users and Groups
,
Users
.
3.
Right-click on
Agilent
, click on
Set Password
, and complete these steps in the window:
a. Type the password obtained from Agilent as the
Password
. Refer to
“Set Up the Agilent
Administrator Password” on page 7-67
for information if you have not yet obtained this password.
b. Type the password again in the
Confirm password
block.
c. Click
OK
.
4.
Right-click on
Agilent
, click on
Properties
, and complete these steps in the window:
a. Unselect
User must change password at next logon
(if not already unselected).
b. Unselect
User cannot change password
(if not already unselected).
c. Select
Password never expires
(if not already selected).
d. Click
OK
.
5. Close the
Computer Management
window.
Install Backup Files onto the New Hard Disk Drive
The files that were previously saved onto a USB memory drive must now be installed onto the new HDD. The
network analyzer must be powered up and operating.
1. Insert the USB memory drive containing the backup files into a USB port.
2. Click on
START
. (It may be necessary to exit the
Network Analyzer
application.)
3. Point to
Programs
,
Accessories
, and then open
Windows
Explorer
.
4. Navigate to C:\Program Files\Agilent\Network Analyzer.
5. Copy each of the backup files from the USB memory drive to the HDD.
Содержание E8362C
Страница 11: ...Service Guide E8364 90038 1 1 1 Safety and Regulatory Information ...
Страница 21: ...Service Guide E8364 90038 2 1 2 General Product Information ...
Страница 33: ...Service Guide E8364 90038 3 1 3 Tests and Adjustments ...
Страница 81: ...Service Guide E8364 90038 4 1 4 Troubleshooting ...
Страница 137: ...Service Guide E8364 90038 5 1 5 Theory of Operation ...
Страница 171: ...Service Guide E8364 90038 6 1 6 Replaceable Parts ...
Страница 257: ...Service Guide E8364 90038 7 1 7 Repair and Replacement Procedures ...
Страница 341: ...Service Guide E8364 90038 8 1 8 General Purpose Maintenance Procedures ...
Страница 382: ...Index 20 Service Guide E8364 90038 Index ...