Chapter 5
237
Language Reference
DISPlay Subsystem
Window Annotation
:DISPlay:WINDow:ANNotation[:ALL] OFF|ON|0|1
:DISPlay:WINDow:ANNotation[:ALL]?
Turns the screen annotation on or off for all windows.
Factory Preset
and *RST:
On
Front Panel
Access:
Display, Preferences, Annotation On Off
Trace Graticule Display
:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe] OFF|ON|0|1
:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe]?
Turns the graticule on or off.
Factory Preset
and *RST:
On
Front Panel
Access:
Display, Preferences, Graticule On Off
Trace X-Axis Scale Offset
:DISPlay:WINDow:TRACe:X[:SCALe]:OFFSet <freq>
:DISPlay:WINDow:TRACe:X[:SCALe]:OFFSet?
Specifies the frequency offset for all frequency readouts such as center frequency,
except that it does not affect marker count.
Factory Preset
and *RST:
0 Hz
Range:
–500 THz to 500 THz
Default Unit:
Hz
History:
Prior to firmware revision A.06.00, the lower range is –3 GHz.
Remarks:
Frequency offset is not available when frequency scale type is
Log (
[:SENSe]:SWEep:SPACing LINear|LOGarithmic
).
Front Panel
Access:
FREQUENCY/Channel, Freq Offset
Содержание E7401A