8
Chapter 1
E4428C Analog Signal Generator Overview
Front Panel Overview
9. EXT 1 INPUT Connector
This BNC input connector accepts an input signal for use with AM, FM,
Φ
M, and pulse modulation. The
damage levels are 5 V
rms
and 10 V
p
.
On signal generators with Option 1EM, this input is relocated to a female BNC connector on the rear panel.
10. EXT 2 INPUT Connector
This BNC input connector accepts an input signal for use with AM, FM,
Φ
M, and pulse modulation. The
damage levels are 5 V
rms
and 10 V
p
.
On signal generators with Option 1EM, this input is relocated to a female BNC connector on the rear panel.
11. Help Key
Press this key to display a short description of any hardkey or softkey. There are two help modes available:
single and continuous. Single mode is the factory preset condition. To toggle between single and continuous
mode, press
Utility
>
Instrument Info/Help Mode
>
Help Mode Single Cont
.
12. Trigger Key
This hardkey initiates an immediate trigger event for a function such as a list or step sweep. The trigger
mode must be set to
Trigger Key
prior to initiating a trigger event with this hardkey.
13. LF OUTPUT Connector
This BNC connector is the output for modulation signals generated by the low frequency (LF) source
function generator. This output is capable of driving 3 V
p
(nominal) into a 50
Ω
load.
On signal generators with Option 1EM, this output is relocated to a rear-panel female BNC connector.
AM, FM,
Φ
M
±
1 V
p
produces the indicated deviation or depth. When using the AC coupled input selection with a signal that
has a peak input voltage that differs from 1 V
p
by more than 3%, the signal generator displays HI/LO annunciator.
Pulse Modulation
+
1 V is on and 0 V is off
AM, FM,
Φ
M
±
1 V
p
produces the indicated deviation or depth. When using the AC coupled input selection with a signal that
has a peak input voltage that differs from 1 V
p
by more than 3%, the signal generator displays HI/LO annunciator.
Pulse Modulation
+
1 V is on and 0 V is off
Single Mode
The help text is provided for the next key you press without activating the key’s function. Any key pressed
afterward exits the help mode and activates the key’s function
Continuous Mode
The help text is provided for each subsequent key press until you either press the
Help
key again, or change to
single mode. In continuous mode, pressing the
Help
key also activates the key’s function (except for the
Preset
key).
Содержание E4428C
Страница 22: ...Contents xxii ...
Страница 107: ...Chapter 3 83 Basic Operation Using Security Functions Figure 3 6 ESG Screen with Secure Display Activated ...
Страница 182: ...158 Chapter 4 Basic Digital Operation Using Waveform Clipping Figure 4 22 Rectangular Clipping ...
Страница 183: ...Chapter 4 159 Basic Digital Operation Using Waveform Clipping Figure 4 23 Reduction of Peak to Average Power ...
Страница 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Страница 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Страница 229: ...205 6 Analog Modulation ...
Страница 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Страница 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Страница 287: ...263 9 BERT This feature is available only in E4438C ESG Vector Signal Generators with Option 001 601or 002 602 ...
Страница 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Страница 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Страница 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Страница 454: ...430 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Concepts Figure 15 9 Uplink Channel Structure ...
Страница 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Страница 667: ...643 18 Troubleshooting ...
Страница 700: ...Index 676 Index ...