Performance Tests
Output Standing Wave Ratio (SWR) Test
Agilent E4418B/E4419B Service Guide
2-29
Table 2-7: Output SWR Result
Max
Actual Result
1.06
1
1.08
2
1. For instruments with the 50MHz 1mW output
(power ref) connector mounted on the front panel.
2. For instruments with the 50MHz 1mW output
(power ref) connector mounted on the rear panel. This
applies to models E4418B and E4419B with option 003
fitted.