Performance Tests
Instrument Accuracy Test
2-8
Agilent E4418B/E4419B Service Guide
Test Setup
Figure 2-2: Instrument Accuracy Test Setup
Procedure
The following procedure should be performed for the Instrument Accuracy
Test.
Note
The procedure details the key presses required on the
Agilent E4418B. For the Agilent E4419B the equivalent key
presses should be performed on both channels.
1.
Connect the equipment as shown in Figure 2-2.
2.
Switch the power meter on.
3.
Press
then
.
4.
Press
,
for a reading in watts.
5.
Set the range calibrator as follows:
RANGE............................................ 3
µ
W
POLARITY ...................................... NORMAL
FUNCTION..................................... STANDBY
LINE ............................................... ON
Agilent 11730A
POWER
METER
CHANNEL A
Power Meter
Range
Calibrator
Preset
Local
Confirm
dBm/W
W
4402serv.book Page 8 Tuesday, October 14, 2003 3:18 PM
Содержание E4418B
Страница 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Страница 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Страница 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Страница 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Страница 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Страница 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Страница 154: ...Index 4 Agilent E4418B E4419B Service Guide 4402serv book Page 4 Tuesday October 14 2003 3 18 PM ...