Chapter 3 Front-Panel Operation and Features
Disabling the Output
56
Disabling the Output
The output of the power supply can be disabled or enabled from the front panel.
• When the power supply is in the “Off” state, the
OFF
annunciator turns on
and the output is disabled. The
OFF
annunciator turns off when the power
supply returns to the “On” state. When the output is disabled, the voltage
value is 0 volts and the current value is 0.02 amps. This gives a zero output
voltage without actually disconnecting the output.
• The output state is stored in
volatile
memory; the output is always disabled
when power has been off or after a remote interface reset.
• While the output is disabled, the control
knob
is locked to prevent from any
unwanted changes occurring. But the other front panel keys are working.
• To lock the control knob, move the flashing digit to the right or left using
or
resolution selection keys until the flashing digit disappears.
To see or check the changes while the output is disabled, press
before
returning to meter mode.
•
Front-panel operation:
output Off
•
Remote interface operation:
OUTP {OFF|ON}
Disconnecting the Output Using an External Relay
To disconnect the output, an external relay must be connected between the
output and the load. A TTL signal of either low true or high true is provided to
control an external relay. This signal can only be controlled with the remote
command
OUTPut:RELay {OFF|ON}
. The TTL output is available on the RS-
232 connection pin 1 and pin 0.
When the
OUTPut:RELay
state is “ON”, the TTL output of pin 1 is high
(4.5 V) and pin 9 is low (0.5 V). The levels are reversed when the
OUTPut:RELay
state is “OFF”. The TTL output of pin 1 or pin 9 of the RS-232
connector is available only after installing two jumpers (JP1 and JP2) inside
the power supply. See the
Service Information
to locate them.
N o t e
Do not use the RS-232 interface if you have configured the power supply to output
relay control signals. Internal components on the RS-232 circuitry may be damaged.
<
>
Display
Limit
On/Off
Output
Содержание E364XA Series
Страница 9: ...8...
Страница 15: ...14 Contents Contents...
Страница 16: ...1 Quick Start...
Страница 26: ...2 General Information...
Страница 38: ...3 Front Panel Operation and Features...
Страница 70: ...4 Remote Interface Reference...
Страница 122: ...5 Error Messages...
Страница 133: ...Chapter 5 Error Messages Calibration Errors 132...
Страница 134: ...6 Application Programs...
Страница 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Страница 146: ...7 Tutorial...
Страница 157: ...Chapter 7 Tutorial Remote Programming 156...
Страница 158: ...8 Specifications...
Страница 165: ...Chapter 8 Specifications Supplemental Characteristics 164 Figure 8 2 Dimensions for Rack mounting...
Страница 166: ...Appendix Service Information...
Страница 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Страница 207: ......
Страница 208: ......
Страница 209: ......
Страница 210: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......