60 Register-Based Programming
Appendix B
Example: Using the Register and Pin-out Maps
As an example of using the register information, assume you want to use the
E1474 to close a remote 5V relay, K1 (see Figure B-2), connected to the
RMD 2 connector (see Figure 2-3). First, change J400 on the driver board
to the +5V position (see Figure 2-4).
NOTE
All relays connected through J400 must now be 5 volt relays.
From the Register maps, Channel 01 on RMD2 is Bit 01. To close channel
01 (relay K1, Figure B-2) on RMD 2 [pin 19 of P1 (corresponds to pin 42 of
P400) needs to be at 0 volts to turn on relay K1], send:
CLOSE (@010101)
or
VXI: WRIT ladd,20,2
where
ladd
is the logical address of the module.