Configuring the RF Multiplexers 17
Chapter 2
Chapter 2
Configuring the RF Multiplexers
Using This Chapter
This chapter shows how to connect external wiring to the RF Multiplexer and
Expander connectors and how to configure the modules for operation.
•
Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
•
Setting the Address Switch . . . . . . . . . . . . . . . . . . . . . . . . . . 18
•
Selecting the Interrupt Priority . . . . . . . . . . . . . . . . . . . . . . . . 19
•
Expanding the RF Multiplexer . . . . . . . . . . . . . . . . . . . . . . . . 20
•
Selecting the Relay Drive Level . . . . . . . . . . . . . . . . . . . . . . . 22
•
Connecting User Inputs to E1472A/73A . . . . . . . . . . . . . . . . 24
•
Connecting User Inputs to E1474A/75A . . . . . . . . . . . . . . . . 25
•
Cables and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Warnings and Cautions
WARNING
SHOCK HAZARD. Only service-trained personnel who are
aware of the hazards involved should install, remove, or
configure the RF Multiplexer. Before you remove any installed
module, disconnect AC power from the mainframe and from
other modules that may be connected to the RF Multiplexer.
CHANNEL WIRING INSULATION. All channels that have a
common connection must be insulated so that the user is
protected from electrical shock in the event that two or more
channels are connected together. This means wiring for all
channels must be insulated as though each channel carries the
voltage of the highest voltage channel.
CAUTION
MAXIMUM POWER.
The maximum power that can be applied to any
SMB connector is 24 W (24 VA). The maximum voltage that can be applied
to any SMB connector is 42V peak. The maximum current that can be
applied to any SMB connector is 1A DC/AC RMS.
STATIC ELECTRICITY.
Static electricity is a major cause of component
failure. To prevent damage to the electrical components in the RF
Multiplexer, observe anti-static techniques whenever removing a module
from the mainframe or whenever working on a module.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание E1472A
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 9: ...8 Notes Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 69: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...