In this program, the command module at logical address 64 has a
primary GPIB address of 08. It has a servant pointer setting of 32,
thus its servant area is from logical address 65 to logical address 96.
If the Agilent E1411B multimeter has a logical address of 80, its
secondary GPIB address is 10. Thus, when programming this
multimeter, its GPIB address is:
OUTPUT 70810;"....
When programming this command module, its GPIB address is:
OUTPUT 70800;"...
•
The following errors are associated with the Interrupt Line
Allocation table:
ERROR 24: INTERRUPT LINE UNAVAILABLE
This error occurs when an interrupt line assigned by the
user-defined interrupt line allocation table is not available.
Either the line has already been assigned or has been reserved.
This error also occurs if the line being assigned to an interrupter
is not handled by the interrupter’s commander.
ERROR 25: INVALID UDEF HANDLER
This error occurs when the logical address specified in the
user-defined interrupt line allocation table for the interrupt
handler (Handler Laddr) is a device that is not a valid interrupt
handler.
ERROR 26: INVALID UDEF INTERRUPTER
This error occurs when the logical address specified in the
user-defined interrupt line allocation table for the interrupter
(Intr # Laddr) is a device that is not a valid interrupter.
ERROR 41: INVALID UDEF INTR TABLE
This error occurs when the user-defined interrupt line allocation
table is not true (valid flag does not equal 1).
ERROR 42: INVALID UDEF INTR TABLE DATA
This error occurs when the user-defined interrupt line allocation
table has invalid data; the number of records and/or the interrupt
line specified is less than 1 or greater than 7, or there is an
invalid interrupt handler and/or interrupter logical address (valid
addresses are 0 to 255).
•
The interrupts assigned by the interrupt line allocation table are used by
the system until
DIAGnostic:BOOT:COLD
or
VXI:CONFigure:ITABle 0
is executed.
Chapter 2
Configuring the Agilent E1406A Command Module 59
Содержание E1406A
Страница 2: ......
Страница 14: ...Notes 12 Agilent E1406A Command Module User s Manual ...
Страница 15: ...Notes Agilent E1406A Command Module User s Manual 13 ...
Страница 16: ...Notes 14 Agilent E1406A Command Module User s Manual ...
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......
Страница 101: ......
Страница 102: ......
Страница 120: ...Notes 118 Triggering and System Status Chapter 4 ...
Страница 179: ... DIAG BOOT COLD will set PARity to NONE Chapter 5 Agilent E1406A Command Reference 177 ...
Страница 240: ...Notes 238 Agilent E1406A Command Reference Chapter 5 ...
Страница 252: ...Notes 248 Agilent E1406A Specifications and General Information Appendix A ...