Display
Explanation
1
Testing ROM
Testing 512 KB RAM
Passed
Testing CPU
CPU Self Test Passed
Non-volatile Ram Contents Lost
GPIB address: 09
Talk/Listen
command module ladd = 0
command module servant area = 255
The Agilent E1406A operating system
performs a series of self-tests and clears its
volatile RAM. The command module’s
GPIB address, logical address, and servant
area (based on the switch settings) are
reported.
2 Command Module VMEbus timeout -- ENABLED
The resource manager identifies the status
of the command module VMEbus timeout.
This must be ENABLED for systems without
VXIbus extenders (Agilent E1406A
Command Module GPIB switch #5 = 0).
3 Searching for static devices in mainframe 0
SC device at ladd 0 in slot 0
SC device at ladd 8 in slot ?
SC device in ladd 16 in slot 8
Searching for dynamic devices in mainframe 0
DC device in slot 3 moved to ladd 24, block size = 1
The resource manager identifies all
statically configured modules, and then
locates and configures all dynamically
configurable modules.
4 Searching for pseudo devices
Pseudo devices are instruments such as
IBASIC.
5 Configuring Commander/Servant hierarchy
ladd = 0, cmdr ladd = -1
ladd = 8, cmdr ladd = 0
ladd = 16, cmdr ladd = 0
ladd = 24, cmdr ladd = 0
ladd = 32, cmdr ladd = 24
ladd = 64, cmdr ladd = 24
Validating Commander/Servant hierarchy
Commander ladd 24 granted device ladd 32
Commander ladd 24 granted device ladd 64
The resource manager establishes the
VXIbus system’s commander/servant
hierarchies based on the commander’s
servant area and the servant’s logical
address.
6 Mapping A24 Memory
ladd 0, offset = 00200000H, size = 131,072 (bytes)
ladd 24, offset = 00220000H, size = 131,072 (bytes)
ladd 64, offset = 00240000H, size = 131,072 (bytes)
Mapping A32 memory in mainframe 0
The resource manager allocates A24
addresses to access the memory located on
the modules at logical addresses 0, 24, and
64. The offset is specified in hexadecimal
and the size is specified in bytes. In this
system, there are no A32 devices.
7 Configuring VME interrupts
VME interrupt line 1 assigned to ladd 0, handler ID 1
VME interrupt line 2 assigned to ladd 24, handler ID 1
VME interrupt line 3 assigned to ladd 64, handler ID 1
VME interrupt line 4 - no handler assigned
VME interrupt line 5 - no handler assigned
VME interrupt line 6 - no handler assigned
VME interrupt line 7 - no handler assigned
The resource manager allocates interrupt
lines to itself and to the other interrupt
handlers in the system.
8 SYSTEM INSTALLED AT SECONDARY ADDR 0
VOLTMTR INSTALLED AT SECONDARY ADDR 1
SWITCH INSTALLED AT SECONDARY ADDR 2
MBinstr INSTALLED AT SECONDARY ADDR 3
SYSTEM instrument started
BNO issued to ladd 24, BNO response = FFFE
Opening GPIB access for message-based device at sec addr 03
The resource manager identifies the
secondary GPIB addresses used in the
system, starts the system instrument (i.e.,
command module), issues the Begin Normal
Operation (BNO) command to its direct
message based servant, and opens GPIB
access to the module at secondary GPIB
address 03.
Figure 2-3. Resource Manager Configuration Without Extenders
Sequence
46 Configuring the Agilent E1406A Command Module
Chapter 2
Содержание E1406A
Страница 2: ......
Страница 14: ...Notes 12 Agilent E1406A Command Module User s Manual ...
Страница 15: ...Notes Agilent E1406A Command Module User s Manual 13 ...
Страница 16: ...Notes 14 Agilent E1406A Command Module User s Manual ...
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......
Страница 101: ......
Страница 102: ......
Страница 120: ...Notes 118 Triggering and System Status Chapter 4 ...
Страница 179: ... DIAG BOOT COLD will set PARity to NONE Chapter 5 Agilent E1406A Command Reference 177 ...
Страница 240: ...Notes 238 Agilent E1406A Command Reference Chapter 5 ...
Страница 252: ...Notes 248 Agilent E1406A Specifications and General Information Appendix A ...