Step 3: Set the memory module address switch(es) according to Table 2-1.
For example, in the figure above, the four Most Significant Bit (MSB) switches are shown in the "1" position,
the next bit is shown in the "0" position. The three Least Significant bit switches (shown in the "1" position)
are not used and may be in either the "1" or "0" position. Only the four Most Significant Bit address switches
need to be set for the 1 MByte modules; or five Most Significant Bit address switches for the 512 kByte
modules. Note: Ignore any numbering on the switch bank. Use the figure above to properly orient the address
switch.
Table 2-1. Memory Module Address Switch Settings
Module(s) being
installed:
Set one module’s address to:
Set the other module’s address to:
One
1
MByte
Module
MSB LSB
1 1 1 1 x x x x
Two
1MByte
Modules
MSB LSB
1 1 1 1 x x x x
MSB LSB
1 1 1 0 x x x x
("x" signifies a “ Don’t Care” state)
Installing Memory into an Agilent E1306A Command Module 2-3