ADM Flow Meter
Notes, Cautions, and Warnings
8
Agilent CrossLab CS ADM Flow Meter Operation Manual
Notes, Cautions, and Warnings
NOTE
•
The ADM flow meter cartridge is a calibrated device.
•
Using the USB port to supply power does not recharge the
installed batteries.
•
The ADM flow meter can also be powered by
rechargeable batteries, but the battery indicator may not
display correctly.
CAUTION
•
Do not use the ADM flow meter in temperatures outside
of its specified operating range.
•
The ADM flow meter should not be misused for
other purposes.
•
The measurement calibration can be affected if the
instrument is exposed to temperatures above 50 °C for
extended periods of time.
•
Please inspect the temperature label attached to the
inside of cartridge packaging to verify that your cartridge
has not been damaged in shipment. A gray circle is
acceptable, a black circle indicates that the cartridge
may have been exposed to an extreme temperature.
•
Please contact Agilent Technical Support if your
cartridge is delivered with a temperature indicator
that is black.
•
Do not use the flow meter to measure corrosive gases.
WARNING
•
Use only a 5 V DC, 0.5 A USB power source and the
supplied USB cable to run this device.
•
Do not use the ADM flow meter as safety equipment.